3.3.5 44°C Program (Fig1.3)
This mode is the pre-programmed default 44°C incubation cycle. In this mode the incubator will
run a 24 hour incubation cycle at 44°C. The timer can be altered to any setting between 0 and 24
hours. Adjusting the incubation time is described in section 3.3.8.
Pressing and holding the middle switch for 5 seconds will start the incubation cycle and the screen
will change to the following:
The asterisk preceding the temperature will switch on and off indicating the status of the heater.
To stop or pause the incubation cycle press and hold the middle switch again for 5 seconds. The
display will show the following:
Pressing and holding the middle switch will restart the cycle or pressing the left switch will return
to the menu function screen.
20.9°C 20.9°C
->44°C MEM
OFF
*20.9°C 20.9°C
24:00 RUN! 24:00
*43.9°C 20.9°C
18:00 MEM 24:00
Содержание AquaSafe WSL50
Страница 1: ...AquaSafe WSL50 Water Safety Laboratory Instruction Manual...
Страница 4: ...Section 2 Kit Contents 1 2 3 4 5 5 Under Accessories Box 6 7 8 a b c d e f g h i j k k l b b a b...
Страница 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual...
Страница 28: ...Section 4 AquaSafe Membrane Filtration Instruction Manual Contents...
Страница 44: ...Section 5 AquaSafe Turbidity Tube Instruction Manual...
Страница 47: ...Section 6 AquaSafe Block Tester Instruction Manual for WSL Range...