Service instructions TR Comfortline
How to calibrate/initialize the lifting columns
From July 2013:
Following calibration instruction is for bathtubs
supplied from July 1
st
, 2013:
- Press the UP and DOWN button simultaneously on the hand control to
calibrate the lifting columns.
- After approximately 10-12 seconds you will hear a “click sound” from the
control box and after approximately 5 seconds more, you will hear an
audible tone from the control box.
- Keep pressing the UP and DOWN button and the lifting columns will
start to move to the highest position. At the highest position you
will hear an audible tone and the system is ready to be used.
Note:
If the tub is not at the lowest position when you start to calibrate, the lifting
columns will move to the lowest position first and then move to the highest
position. Keep both the UP and DOWN button pressed during the
calibration.
Before June 2013:
Old control box used until June 2013:
Please press the DOWN key on the hand control to initialize the
lifting columns. When you reach the lowest position, keep the DOWN
button pressed for 5-8 seconds. When an audible tone is heard
the system is ready to be used.