Page 34
14. Initialising the WP-84
If the calibration settings of the
WP-84
exceed the allowable limits, the unit may
need to be initialised to factory default values. This action may be required if the
sensor is replaced.
To initialise the
WP-84
…
1.
Switch the
WP-84
off by pressing the
key.
2.
Press and hold the
key while switching the
WP-84
back on.
3.
The following messages should be displayed…
Initialized
MUST ReCalibrate
then…
WP84s V2.0 S1234
Cond. Sal. Temp.
(The “s” after
WP-84
is shown when the RS232 serial port option is fitted)
4.
The meter then displays Conductivity and Temperature. Note that the decimal
points have been replaced with a “
∗∗
” to indicate that the unit requires re-
calibration.
Note
When the
WP-84
is initialised, the manual k factor selection is re-set to k=10. See
section 13 if you wish to select a k=0.1 sensor.
15. Instrument firmware version number.
If you need to phone or fax TPS for any further technical assistance, the version
number of your
WP-84
firmware may of benefit to us. The version number is
displayed by the
WP-84
at turn-on.