Page 25
9. Automatic Datalogging
The
WP-84
can automatically log records into the Notepad. First the logging
period must be programmed, then automatic logging can be started and stopped as
required.
1.
Select the Program menu (
→
→
F2:Notepad
→
→
F4:Prog.
).
2.
The display should now look like this…
→
→
00
←
←
F
1:
Min
F
2:
Sec
↑↑↓↓
:Period F
3:
Hr
3.
Use the
and
keys to set the period at which the
WP-84
will
automatically log records.
4.
When the logging period has been correctly set, select whether this period is in
minutes, seconds or hours.
Press
to save the period as minutes.
Press
to save the period as seconds.
Press
to save the period as hours.
For example, if the period was set to
05
, followed by
, then the
WP-84
will
automatically log a record every 5 seconds.
5.
If the optional RS232 port is fitted, the
WP-84
will ask if the records are to be
logged into the Notepad, or sent directly to the RS232 port.
Press
to log records into the Notepad (maximum of 150 readings).
Press
to send records directly to the RS232 port.
6.
The automatic logging function is now programmed, and can be started and
stopped as required.
7.
To start automatic logging, press
in normal display mode.
If the
WP-84
is logging into the Notepad, the display will look like this…
2.76mS 25.0
o
c
Log# 1 12:00:00
The log number will increment and the
WP-84
will beep each time a reading is
recorded.