Page 33
13. Selecting k=0.1 or k=10 Sensors
The
WP-84
automatically recognises a k=1.0 sensor. The
WP-84
does not
automatically recognise k=0.1 or k=10 sensors. When a k=0.1 or k=10 sensor is
used, the
WP-84
must be set to the correct k factor before use. The following
procedure describes how to select a k=0.1 or k=10 sensor.
1.
Switch the meter
OFF
with the
key.
2.
Connect the k=0.1 or k=10 sensor.
3.
Press and HOLD the
key while switching the meter back on.
4.
The k factor selection menu is now displayed (only if the k=0.1 or k=10 sensor
is connected). For example…
Select F1:k=.1
k Factor >F2:k=10
The arrow indicates the current selection.
5.
Press
to select a k=0.1 sensor.
Press
to select a k=10 sensor.
Press
to quit buffer selection and retain the current setting.
Notes
1.
The manual k factor selection is kept in memory when the meter is switched
off, even if the battery is removed.
2.
The manual k factor selection is reset to k=10 during initialisation.
3.
The
WP-84
will always automatically recognise a k=1.0 sensor, regardless of
the manual k factor selection.
4.
Calibration settings for k=0.1, k=1.0 and k=10 sensors are
NOT
stored
separately. The
WP-84
requires re-calibration when a new k factor sensor is
connected.
5.
The message “
Connect k=10/0.1
” appears when the
WP-84
cannot
detect a k=10 or k=0.1 sensor connected to the
Sensor
socket. Check that the
sensor is correctly connected. If the message persists, return the k=10 or k=0.1
sensor to the factory for checking.