Page 35
11.3.3 Temperature Compensation
The slope of a pH sensor (section 11.3.2) is affected by temperature. This
effect is compensated for either by using an Automatic Temperature
Compensation (ATC) sensor or by entering the sample temperature
manually. Figure 11-3 shows the slope of a pH sensor at various
temperatures.
-600
-400
-200
0
200
400
600
0
7
14
Electrode
Potential (mV) at
0 oC (54mV/pH)
Electrode
Potential (mV) at
50 oC (64mV/pH)
Electrode
Potential (mV) at
100 oC (74mV/pH)
pH Electrode Response, as a Function of Temperature
E
le
c
tr
o
d
e
R
e
s
p
o
n
s
e
(
m
V
)
pH
Figure 11-3