Page 28
Conductivity, TDS and Salinity Troubleshooting, continued...
Inaccurate
readings, even
when calibration is
successful.
1. Sensor may have a build-
up of dirt or oily material
on sensor wires.
2. Platinum-black coating
has worn off.
Clean sensor, as per the
instructions detailed in
section 11.1.
Sensor requires
replatinisation. Return to the
factory, or see details in
section 11.2.
Readings drift.
1. Sensor may have a build-
up of dirt or oily material
on sensor wires.
Clean sensor, as per the
instructions detailed in
section 11.1.
Readings are low
or near zero.
1. Sensor may have a build-
up of dirt or oily material
on sensor wires.
2. Sensor is not immersed
deeply enough.
3. Sensor is faulty.
Clean sensor, as per the
instructions detailed in
section 11.1.
Immerse sensor at least to
the vent hole in the white
plastic cover.
Return sensor to factory for
repair or replacement.
Display flashes
“ATC” and “LIMIT”
The Temperature is not
within the ATC limits.
Cool/Heat solution before
taking measurements.