Page 21
6. Selecting k=10 or k=0.1 sensor
The
Aqua-CPA
automatically recognises a k=1.0 sensor. The
Aqua-CPA
does not
automatically recognise k=0.1 or k=10 sensors. When a k=0.1
or k=10 sensor is used, the
Aqua-CPA
must be set to the correct k factor
before use. The following procedure describes how to select a k=0.1 or
k=10 sensor.
1. Switch the meter
OFF
.
2. Connect the k=0.1 or k=10 sensor.
3. Press AND HOLD the
key while switching the meter back on.
The k factor selection menu is now displayed (only if the k=0.1 or
k=10 sensor is connected)…
k=10
↑
↑
↑
↑↓
↓
↓
↓
k Factor
or
↔
k=0.1
↑
↑
↑
↑↓
↓
↓
↓
k Factor
4. Press the
and
keys to alternate between a k=10 and a k=0.1
sensor.
5. Press the
key to save the selected setting.
Notes
1. The manual k factor selection is kept in memory when the meter is
switched off, even if the battery is removed.
2. The manual k factor selection is reset to k=10 during initialisation.
3. The
Aqua-CPA
will always automatically recognise a k=1.0 sensor,
regardless of the manual k factor selection.
4. Calibration settings for k=0.1, k=1.0 and k=10 sensors are
NOT
stored separately. The
Aqua-CPA
requires re-calibration when a new
k factor sensor is connected.