Page 17
DO NOT
place the sensors directly into the buffer bottle. Discard the
used buffer after use.
9. When the reading has stabilised, press the
key to calibrate. The
Aqua-CPA
will prompt the user to select the parameter if a
combination display is shown. Use the
or
keys to select pH. If
a 1 point calibration has been performed, a “
∗
“ will not be removed
until a full 2 point calibration has been performed.
10. Rinse the pH and Conductivity sensors in distilled water and blot
them dry.
11. Place both sensors into a small sample of pH4.01, pH9.18 or
pH10.01 Buffer, so that the bulb and reference junction are both
covered, as per the diagram in step 8.
DO NOT
place the sensors
directly into the buffer bottle. Discard the used buffer after use.
pH9.18 and pH10.01 buffers are unstable once the bottles have
been opened. Discard immediately after use
.
12. When the reading has stabilised, press the
key to calibrate. The
Aqua-CPA
will prompt the user to select the parameter if a
combination display is shown. Use the
or
keys to select pH.
The “
∗
“ will now be replaced by a decimal point, if calibration was
successful. The
Aqua-CPA
is calibrated for pH and is ready for use.