Page 25
9. Initialising the Aqua-CPA
If the calibration settings of the
Aqua-CPA
exceed the allowable limits,
and the unit cannot be re-calibrated, then it may need to be initialised to
factory default values. This action may be required if an sensor is
replaced.
To initialise the
Aqua-CPA
…
1. Switch the
Aqua-CPA
off.
2. Press AND HOLD the
key while switching the
Aqua-CPA
on.
3. The following messages are now displayed…
Memory & Cal
Reset !
↓
↓
↓
↓
You MUST
Re-Calibrate
↓
↓
↓
↓
TPS AQUA-CPA
V2.0 S1234
4. The meter then displays Conductivity, pH and Temperature. Note that
the decimal points have been replaced with a “
∗
“, to indicate that the
unit requires re-calibration.