STORING DATA
During testing data can be stored for later retrieval.
When data is saved, all data previously in the address will be overwrit-
ten.When saving pressure/draft readings, select an address different from
the one used to save combustion readings.
1. Press the Store Key.
2. “Addr” and an address location from 0 to 99 will flash on the display. Use
the Up/Down Arrow Keys to select the desired location to store the data
and press the Scroll/Enter Key.
To cancel storing information press the
Function Key.
3. The data is stored and the analyzer returns to normal operation.
You have just successfully stored a set of readings which can be reviewed
on screen (see pg 26) or sent to the IR printer (see pg 27)
25
www.
.com