Combustion Analyzer
(Default Start Up Mode)
COMBUSTION TEST PROCEDURE
Note: It is recommended you perform routine general maintenance on your
analyzer to ensure proper function. Please refer to Appendix A for general
maintenance schedule and function tests.
1. Turn the 717 on in fresh air as outlined on page 8. After the initial purge
cycle the 717 will default to combustion analyzer mode and combustion dis-
play 1 will be seen.
•
Combustion Display 1 (Default Start Up Display)
• Displays Oxygen (O2) reading in percent (%)
• Displays Carbon Monoxide (CO) in parts per million (ppm)
2. Connect the Mini Pump Protection Filter assembly and Flue Probe Tubing
complete with In-Line Filter to the Gas Sample Port and the 'K' Type
Thermocouple Plug from the Flue Probe into Thermocouple (T1) Socket. The
GK11M ambient air temperature probe is connected to the (T2) socket. (See
below & page 7)
WARNING
: - Ensure the 'K' type thermocouple probes are inserted into the
sockets correctly (see page 7). The plugs are polarity marked and forcing the
plug into the socket the wrong way may result in damage to the
instrument.
Mini Pump
Protection Filter
Assembly
Ambient Air
Connection (T2)
GK11M Probe
Flue Probe
Connection (T1)
In-line Filter
Assembly
Flue Probe
Tubing