The information about EMC conformity (IEC60601-1-2:2007)
Guidance and manufacturer’s declaration - electromagnetic emissions
The X-Ray FLAT PANEL IMAGER (hereafter called FPI) is intended for use in the electromagnetic
specified below.
The customer or the user of the FPI should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group1
The FPI uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class A
The FPI is suitable for use in all establishments other
than domestic and those directly connected to the
public low-voltage power supply network that supplies
building used for domestic purposes.
Harmonic emissions
IEC61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Not applicable
Note
The FPI is not allowed to use the cables or components differ from the
originally attached.
If the different cables or components are used, they may deteriorate the
performance of electromagnetic emissions.
The limited length of cables for conforming EMC (IEC60601-1-2:2007)
Cables
Cable length (Max.)
Sensor Unit Cable
14 m
GND Cable
3 m
Refer to Product Components and Interface about wiring and connection of these cables.