MAIN CHARACTERISTICS
Image Format:
X-ray Conversion Layer ................... Cesium Iodide (Csl) with Amorphous Silicon (a-Si) Photodiode
Active Area ........................................................................................ 35(H)
×
43(V)cm (14
×
17 inch)
Pixel Matrix ........................................................................................................... 2448(H)
×
2984(V)
Pixel Pitch ............................................................................................................................ 143
μ
m
Cycle Time ............................................................................................................. Shot to Shot 6sec
Performance:
Limiting Resolution .................................................................................................... 3.7 Lp/mm typ.
MTF (2.0 Lp/mm, 70 kVp, 1
×
1) ......................................................................................... 36 % typ.
DQE (DQE (0), Quantum - Limited) ........................................................................................ > 70 %
A/D Conversion......................................................................................................................... 16 bit
Ratings:
Energy Range ............................................................................................................... 40
~
150 kVp
Maximum Entrance Dose (low Gain)
(
Linear Output Range
)
....................................... 4 mR / frame
Interface (Interface Box):
Sensor Unit
UNIT Interface ............................................................................................. Connect to Interface Box
Interface Box
Unit Interface ................................................................................................ Connect to Sensor Unit
Data Output ................................................................. 16 bit Digital Output Ethernet (1000BASE-T)
Command Control .......................................................................................... Ethernet (1000BASE-T)
X-ray Synchronization Control ..............................................................................................
External
Power Input .............................................................................................. AC100 to 240V, 50/60 Hz
Image Acquisition Exposure period: