FDX3543RP
-7-
FPI Drive
EXP_RE
<IN>
EXP_OK <OUT>
EXPOSURE
DATA_STRAGE
Ready
1st image data
2nd image data
Exposure
Read
Exposure
Read
Ref
Ref
FPI ENABLED
1st
1st
②
①
2nd
2nd
C
D
A
B
C
C
D
C
(2) Double Exposure
A: EXP_REQ width
Minimum 1ms
B: Refresh period
For line drive (2984 Line scan) 80 ms
C: EXP period
Set value: see the Image Acquisition
mode table (500 to 2000 ms (by 500 ms step),
standard 500 ms).
D: Read period
For line drive (2984 Line Scan) 240 ms
①:
EXP_REQ
-
EXP_OK period ........................................................................ Maximum 150 ms
②:
Double EXPOSURE cycle period ................... Maximum 1700 ms (case EXPOSURE 500 ms)