• Re-locate the direction or re-arrange the receiving devices.
• Enlarge the distance between equipment.
• Connect this equipment to the socket of other circuits.
In case where the fault cannot be solved by taking the foregoing measures, please stop using it and
consult our after-sales service personnel.
Precautions for EMC
1. Special attention shall be paid to EMC when using medical electronic equipment and such
equipment shall be installed and used according to the EMC information provided in this manual.
2. Portable and mobile radio frequency communication equipment may influence the running of
medical electronic equipment.
3. Cables that can intervene EMC are as below:
In order to maintain the best EMC performance, please adopt the specialized cables.
Name
Type
Length
Remarks
AC power cord
Non-shielded
Fixed length of 3.0m
Parenthesis
USB cable
Shielded
3.0m at most
Not attached
HDMI cable
Shielded
2.7m at most
Not attached
4. Failure to adopt the specified accessories, energy converters and cables may result in the
increase in the electromagnetic radiation quantity or the decline in anti-interference performance
of Kestrel 3100m.
5. Never place Kestrel 3100m on the accessories of other equipment or other equipment. If it is
mandatory, please check if Kestrel 3100m can work well.
Guidance for Electromagnetic Emission and Manufacturer's
Declaration
Kestrel 3100m is especially designed to work in the following specified electromagnetic
environment.
Customers or users shall ensure that Kestrel 3100m is used in the following environment.
1)
Electromagnetic emission
Emission test
Conformity
Electromagnetic environment - guidance
RF emission
GB 4824
1 set
Kestrel 3100m can exert the radio frequency
energy when running its internal functions.
Therefore, its radio frequency radiation level is
quite low, and there is a small chance to
Содержание Kestrel 3100m
Страница 1: ...Kestrel 3100m Fundus Camera Technical and Operation Manual Shanghai Topview Industry Co Ltd...
Страница 4: ......
Страница 11: ...7 2 3 Division of function modules...
Страница 12: ...8...
Страница 13: ...9...
Страница 21: ...17 3 3 Image acquisition module 1 Image monitoring 2 Fundus alignment Manual and auto...
Страница 35: ...31...