22. CHECKS AND ADJUSTMENTS
With this option you can measure collimation error in your instrument so that the instrument can
correct subsequent single face observations. To measure the error, make angular observations
using both faces.
• Perform adjustment in weak sunlight and no scintillation.
PROCEDURE Adjusting
1. Carefully level the instrument.
2. Install a target at a point about 100m in the
horizontal direction from the instrument.
3. Select "Inst. cons." in <Configuration>.
4. Select "Collimation".
5. While the telescope is in Face 1, sight the center
of the target correctly and press
[OK]
. Rotate the
instrument 180°.
22.3
Collimation