background image

4-42

   4 PROGRAM MODE

4.8

LOADER option

The loader option can load Application programs from a PC to the GTS-600 series.
Before loading any application programs, make sure the communication parameters in the instrument 
match the parameters on the PC software. For further information about the parameters, refer to the 

application software manual.

Operating procedure

Operation

Display

1

Press [F6](MORE) key from programs menu 
to get the next page of programs.

[F6]

2

Press [F3](LOADER) key.
If the instrument and the PC are not 
connected the message will be shown as the 
right display. 
If you do not want to load the program, press 
the [F1](ESC) key.

[F3]

When the loading is completed, the display 
will return to the main menu icons.

Programs

F1 BS     p               4/8

F2 STORE  p

F3 REM    p

F4 MLM    p            MORE

Programs              

F1 REP     p              8/8

F2 LAYOUT  p  

F3 LINE    p

F4 LOADER  p           MORE

GTS-600 Application Loader

   Waiting Connection•••

ESC

GTS-600 Application Loader

   

     Connected•••

        

        GTS-600 series

Main menu

Содержание GTS-600 Series

Страница 1: ...ELECTRONIC TOTAL STATION GTS 600 SERIES INSTRUCTION MANUAL GTS 601 GTS 601AF GTS 602 GTS 602AF GTS 603 GTS 603AF GTS 605 GTS 605AF...

Страница 2: ......

Страница 3: ...is locked and the base fixing screws are tightened Guarding the instrument against shocks When transporting the instrument provide some protection to minimize risk of shocks Heavy shocks may cause the...

Страница 4: ...se eye injury or blindness Do not look at the sun through a telescope Laser beams can be dangerous and can cause eye injury s if used incorrectly Never attempt to repair the instrument yourself Only f...

Страница 5: ...xplained in the user manual 6 The manufacturer or its representatives assumes no responsibility for damage caused by wrong movement or action due to connecting with other products CAUTION Use of contr...

Страница 6: ...lowing labels are found on the instruments which describe the GTS 600 series Laser plummet type Precautions and safety information about the laser beam If at any time any of these labels are damaged a...

Страница 7: ...tching Horizontal Angle Right Left 3 2 3 1 3 Measuring from the Required Horizontal Angle 3 2 3 1 4 Vertical Angle Percent Grade Mode 3 3 3 2 Distance Measurement 3 4 3 2 1 Setting of the Atmospheric...

Страница 8: ...tical Cross hair 8 5 8 3 4 Collimation of the Instrument 8 6 8 3 5 Checking Adjusting the Optical Plummet Telescope 8 7 8 3 6 Checking Adjusting the Laser Plummet For Laser Plummet type 8 8 8 4 Adjust...

Страница 9: ...river cleaning brush 1 set 5 Plastic carrying case 1 each 6 Sun shade 1 each 7 Plastic rain cover 1 each 8 Silicon cloth 1each 9 Instruction manual 1 each Make sure that all of the above items are wit...

Страница 10: ...Base Handle fixing screw Sighting collimator Carrying handle Vertical motion clamp 1 Vertical tangent screw 1 2 GTS 601 601AF 602 602AF 2speed way GTS 603 603AF 605 605AF 1speed way Power supply conn...

Страница 11: ...clamp Horizontal tangent screw 1 Instrument center mark Battery locking lever Battery BT 50Q Adjusting screw for circular level Tribrach fixing lever GTS 601 602 2speed way GTS 603 605 1speed way GTS...

Страница 12: ...ng knob Circular level Adjusting screw for circular level Tribrach fixing lever Horizontal motion clamp Instrument center mark Battery BT 50Q Power switch Horizontal tangent screw 1 GTS 601AF 602AF 2s...

Страница 13: ...e Display marks Angle measurement mode V angle 87 55 20 H angle 180 44 12 Distance measurement mode Horizontal angle 87 55 40 Horizontal distance 180 44 12 Relative elevation 12 345m Display Contents...

Страница 14: ...er key Press at the end of inputting values POWER Power key ON OFF of power source Power key is located on the side of the instrument Horizontal distance measuring Coordinate measuring Angle measuring...

Страница 15: ...2 To be stake out measurement mode MEAN F3 Sets the number of N time measurement m ft F4 Switches meter or feet unit Horizontal distance measuring MEAS F1 Horizontal distance measuring starts Switches...

Страница 16: ...ircular graphic display F1 7 The light acceptance quantity level signal level is displayed F2 8 Set the Temperature Pressure Atmospheric Correction Value PPM and Prism Constant Value PSM F3 9 Turn the...

Страница 17: ...e display back light ON OFF When the back light is OFF the light bulb icon is dark To turn the back light ON press the F3 key Press F3 again to turn the back light OFF 4 Reticle illumination ON 1 to 9...

Страница 18: ...Measuring Mode 8 Setting Temperature Pressure Atmospheric correction value PPM Prism constant value PSM Press the F6 key to get to Screen 2 on the display then press the F3 key on screen 2 The tempera...

Страница 19: ...S 600 instruction manual to change the time or to invalidate the function 1 6 Auto Power Off If no key operation is given for the setting time 1 to 99 minutes the power turns off automatically To set...

Страница 20: ...f automatically while the automatic focusing is working 3 The auto focusing may be completed roughly when the contrast with the target and its circumference is low In this case focus the target manual...

Страница 21: ...clearly observed Adjust diopter 4 Press the F6 AF key Auto focusing will start 5 Look into the telescope and confirm whether the focusing is completely done or not If not focus the target with the foc...

Страница 22: ...Power Connection unnecessary if on board Ni MH battery BT 50Q is used See below for connecting the external battery pack Battery pack BT 3Q Power cord PC 5 is used Large capacity battery pack BT 3L Po...

Страница 23: ...ument horizontally by using the Horizontal motion clamp screw and place the plate level parallel with the line connecting leveling screws A and B and then bring the bubble to the center of the plate l...

Страница 24: ...power switch ON Confirm the battery power remaining on the display Replace with charged battery or charge when battery level is low see Section 2 4 Battery Power Remaining Display Power switch key ON...

Страница 25: ...apter 12 POWER SOURCE AND CHARGING 3 The battery power remaining display shows the power level regarding to the measurement mode now operating The safety condition indicated by the battery power remai...

Страница 26: ...NE 8 Application software Loader option LOADER see Chapter 4 PROGRAM MODE COMMUNICATION MODE This mode is used for follows Setting of PROTOCOL Data file in out see Chapter 6 COMMUNICATION MODE ADJUSTM...

Страница 27: ...the standing axis in the X and Y directions For more information about dual axis compensation see Chapter APPENDIX The display of Vertical or Horizontal angle is unstable when instrument is on an unst...

Страница 28: ...is carried out to delete the error by turning in normal and reverse both sides of telescope so far To adjust or reset the above compensation value see Chapter 8 CHECK AND ADJUSTMENT Enable you to stop...

Страница 29: ...you to enter numerals or alphabet letters such as the file name Example Enter HIL 104 to rename the file name Operating procedure Option Display 1 Press F1 Alpha key to be entering alphabet letter mo...

Страница 30: ...e target at the peak of the triangle mark of the sighting collimator Allow a certain space between the sighting collimator and yourself for collimating 3 Focus the target with the focusing knob If par...

Страница 31: ...tches Note It is possible that the R L switching is prohibited R L Rock To set the R L rock see Chapter 7 PARAMETERS SETTING MODE Operating procedure Operation Display 1 Set the required horizontal an...

Страница 32: ...uired Horizontal angle is possible ENT 1 To revise wrong value move cursor with F6 BS key or input from the beginning by F1 EXIT key to correct value 2 With wrong input value for example 70 setting wi...

Страница 33: ...ure the mode displays angle measurement Operating procedure Operation Display 1 Collimate the center of prism 2 Press F1 SD key or F2 HD key 1 2 Example Horizontal distance mode F2 The result are show...

Страница 34: ...F2 HD key 2 Press F6 key to get the function page as 2 F2 F6 3 Press F3 MEAN key F3 4 Input the setting the number of times and press ENT key 1 Example 4 times N times measurement starts F4 ENT Operat...

Страница 35: ...e measuring after the measurement in held To return to the continuous measuring press F1 MEAS key twice To return to the angle measuring mode press F3 VH key V 90 10 20 HR 120 30 40 PSM 0 0 HD 54 321...

Страница 36: ...es in shorter time than in fine mode Use this mode for the objects which may be slightly unstable Measurement time approx 0 7 seconds The unit to be displayed is 1mm 0 005ft Operating procedure Operat...

Страница 37: ...F6 2 Press F2 S O key and press ENT key 3 Enter the relative elevation for stake out and press ENT key The measuring starts F2 ENT Enter value ENT 4 Collimate the target Prism The difference between t...

Страница 38: ...inates of the occupied point after turning the power off Resume mode OFF Refer to Chapter 7 PARAMETERS SETTING MODE Confirm the angle measurement mode Operating procedure Operation Display 1 Press F3...

Страница 39: ...t new data and press ENT key 1 Measuring starts N coord ENT E coord ENT Z coord ENT 1 To cancel the setting press F1 EXIT key Setting occ point N 0 0000 E 0 0000 Z 0 0000 EXIT BS Complete N E PSM 0 0...

Страница 40: ...coordinate measurement mode to get the function as in page 2 F3 F6 3 Press F2 HT key Previous data will be shown F2 4 Input instrument height and press ENT key 1 Inst HT ENT 5 Input prism height and p...

Страница 41: ...stance Relative elevation z Coordinates of the center of the prism originated from the center point of the instrument n e z Coordinates of unknown point N1 E1 Z1 N1 N0 n E1 E0 e Z1 Z0 Inst h z P h Con...

Страница 42: ...not set 2 Refer to Section 3 1 3 Measuring from the Required Horizontal Angle or Section 4 1 Setting a Direction Angle for Backsight Orientation 3 Pressing F1 MEAS key the measurement mode Continuous...

Страница 43: ...splay 1 Operate the data collector to measure the distance Measurement will be started 2 The result will be shown and transferred to the Data Collector 3 The mode will automatically return to the dist...

Страница 44: ...Measuring will continue at this time F1 3 Press F5 YES key The measurement will start F5 After the measurement the result will be hold then recorded The screen will return to previous display V 90 10...

Страница 45: ...anglefor backsight orientation BS 2 Retaining a coordinate STORE NEZ 3 Remote elevation measurement REM 4 Missing line measurement MLM 5 Repetition angle measurement REP 6 Layout LAYOUT 7 Line measure...

Страница 46: ...RS SETTING MODE The program does not store the backsight coordinate values in memory Example Backsight point A N coordinate 54 321m E coordinate 12 345m Operating procedure Operation Display 1 Press F...

Страница 47: ...preset coordinates will be used for this program Set the the coordinate value of instrument point P0 and set the direction angle from instrument point P0 toward known point A Operating procedure Opera...

Страница 48: ...Press F2 STORE key F2 9 Press F2 Recall NEZ key 10 Collimate P0 the former instrument point F2 Collimate P0 11 Press F5 YES key F5 The coordinates at P1 and direction angle toward P0 are set The displ...

Страница 49: ...which the vertical angle is established In both procedures the reference point should be perpendicular to the remote object 1 With prism height h input Example h 1 5m Operating procedure Operation Dis...

Страница 50: ...nu press F1 EXIT key Operating procedure Operation Display 1 Press F3 REM key F3 2 Press F2 NO key F2 3 Collimate prism 4 Press F1 MEAS key Measuring starts Collimate P F1 Horizontal distance HD betwe...

Страница 51: ...t G will be decided 1 F6 8 Collimate target K Vertical distance VD will be shown 2 Collimate K 1 To return to procedure 3 press F2 HD key To return to procedure 4 press F3 V key 2 To return to main me...

Страница 52: ...re of 2 A B B C mode is completely same as MLM 1 mode Operating procedure Operation Display 1 Press F4 MLM key F4 2 Press F1 A B A C key F1 3 Collimate prism A and press F1 MEAS key Horizontal distanc...

Страница 53: ...prism C will be shown Collimate C F1 9 Press F6 SET key The horizontal distance dHD relative elevation dVD and slope distance between prism A and C F6 10 To measure the distance between points A and D...

Страница 54: ...using the horizontal motion clamp and the horizontal tangent screw Collimate B 6 Press F6 HOLD key F6 7 Recollimte the first target A using the horizontal motion clamp and the horizontal tangent scre...

Страница 55: ...mber of repetitions Quadrupled angle Horizontal angle can be accumulated up to 3600 00 00 minimum reading horizontal angle right or 3600 00 00 minimum reading horizontal angle left In case of 5 second...

Страница 56: ...ate values for the side shot or resection routine If a job name is not created when doing the side shot or resection routine the GTS 600 series will automatically create a default job identifies as 1...

Страница 57: ...s key F4 2 In the Options menu select F1 Create or Selection a Job F1 3 The jobs that are in memory will show on the screen If there are more then four jobs in memory the F6 MORE will scroll to the ot...

Страница 58: ...y The view memory routine shows all the jobs stored in memory This routine also allows the user to select another job in memory There can be multiple pages of jobs The F6 MORE key scrolls to each page...

Страница 59: ...press F6 to back space To Exit Press F1 3 The next screen allows the user to continue erasing the record or to cancel the routine without erasing the record Press F5 to erase the record or F6 not to e...

Страница 60: ...the name Refer to section 2 9 for Alpha and Numeric input Select Number 3 The next prompt screen is to input the point number The record number is shown in the upper left corner of the display If you...

Страница 61: ...ption will erase all the jobs in memory Transfer Jobs The protocol default parameters are highlighted on the GTS 600 series The Transfer Jobs option will send and receive jobs to and from the PC In th...

Страница 62: ...isfied with the protocol parameters press F1 to store the new parameters If you forget to press F1 to set the new parameter the selection you have made will not be stored and the previous setting will...

Страница 63: ...the receive job option and allows for another job to be received Pressing F5 YES will add the job received from the PC to memory without erasing the current file Pressing F6 will abort the receive op...

Страница 64: ...page 2 F6 2 Press F1 Transfer Jobs 3 To send a job press F2 4 Get the PC software ready to receive the job When the PC is ready press F5 YES to send the job If you press F6 NO the software will return...

Страница 65: ...ssing the ENT key the Job Manager menu appears To make sure the job name was changed go to the View Manager option to see the new job name ENT Operating procedure Operation Display 1 At the Job Manage...

Страница 66: ...key to move the cursor from right or left and the F1 key is to exit back to the Coordinate data main menu Press the ENT key to store each field of data After entering the elevation data the point num...

Страница 67: ...ress the ENT key after each entry After entering in the elevation Z and pressing the ENT key the job name point number and coordinates are stored in memory ENT 7 The point number display appears and t...

Страница 68: ...nstruction below for the Search Data View Jobs option Operating procedure Operation Display 1 From the options layout menu press F3 for the Coordinate data option 2 Press F2 for the Search View Jobs F...

Страница 69: ...n memory are shown on the display The job located a t F1 is highlighted indicating the current job If you would like to select another job press keys F2 thorough F4 To view more jobs in memory press t...

Страница 70: ...edure Operation Display 1 Press F3 for the New Point option F3 2 Press F1 to collect side shot F1 3 This screen allows the user to set a direction angle to the backsight or you can skip this if the di...

Страница 71: ...trument height and press the ENT key F5 Type Inst Height 8 This screen is to input the side shot point number Type in the point number and press the ENT key ENT 9 Type in the prism height and press th...

Страница 72: ...r the data is stored the screen to type in another side shot point number appears step 8 Point number increments 1 Type Shot Point Number Note Note In the Search feature a line is under the record num...

Страница 73: ...compute a resection point 3 This screen allows the user to type in a point number for the new resection point After typing in the print number press the ENT key to continue F2 4 Input the instrument...

Страница 74: ...measure and show the horizontal angle horizontal distance and vertical distance F5 13 After the instrument is complete measuring the screen showing the residual error appears Press F5 to continue with...

Страница 75: ...4 for Grid Factor F4 2 To modify the grid factor press F5 If you select F6 NO the program returns to the Coordinate data menu F5 Type Elevation 3 Type in the elevation and press ENT ENT 4 Next type in...

Страница 76: ...actor acceptable press the ESC key or F6 NO ENT 6 After pressing ESC the coordinate data menu appears ESC Grid Factor 1 000000 Modify YES NO Coordinate Data F1 Input Coordinate data F2 Search Data Vie...

Страница 77: ...ey to advance to the next screen F6 3 Press the F2 LAYOUT key The version number will appear on this screen for a few seconds If a Job was created the information screen about the job will appear F2 N...

Страница 78: ...tored in memory after input F1 Type Coord F6 6 a The next display will prompt for the backsight point number The point number can be alpha or numeric If the point number and its coordinates are stored...

Страница 79: ...ENT Type Point Number Type Coord ENT 10 Type in the prism height for the layout point Type Prism Height 11 The angle and distance to the layout point is shown on the display From the backsight point t...

Страница 80: ...tion has distance features for the instrument person to pass along instructions to the rod person to layout the point One option shows the distance to either move BACK towards the instrument or GO awa...

Страница 81: ...person The distance instructions are to move BACK towards the instrument GO away from the instrument and to move RIGHT or LEFT to get back online with the layout point The RIGHT and LEFT instructions...

Страница 82: ...e of information is calculated using the rod height previously entered F5 3 When the measuring point is within 5mm from standard layout point the word KEEP and or signs are displayed Function of GUIDA...

Страница 83: ...m the instrument to prism B and the distance along the base line and the screen will display the vertical distance from prism A to that point the horizontal distance from that point Additionally the v...

Страница 84: ...Horizontal distance is displayed Collimate B F1 8 Press F6 SET key and horizontal distance will be recorded F6 9 Sight line point L Measured data to the line point L is displayed VD Vertical distance...

Страница 85: ...d point G Sight G 12 Press F6 SET key line height LH and horizontal distance Off are displayed F6 To finish the measurement press the F1 EXIT or ESC key To return to operation procedure 9 press the F2...

Страница 86: ...n Display 1 Press F6 MORE key from programs menu to get the next page of programs F6 2 Press F3 LOADER key If the instrument and the PC are not connected the message will be shown as the right display...

Страница 87: ...from Main Menu Icons 1 Press F6 File key Each File status File name File name extension Used memory capacity Date are shown F6 Prog Std Mem Com Adj Para 2001 07 10 15 30 40 MEMORY MANAGE MODE The foll...

Страница 88: ...rotected Operating procedure Operation Display The memory capacity and the remaining memory capacity are shown by pressing F3 Mem from Main Menu Icons 1 Select a file using F5 key or F6 key 2 Press F1...

Страница 89: ...ration Display The memory capacity and the remaining memory capacity are shown by pressing F3 Mem from Main Menu Icons 1 Select a file using F5 key or F6 key 2 Press F2 Ren key Select a file F2 3 Ente...

Страница 90: ...key or F6 key 2 Press F3 Del key Select a file F3 3 Confirm the file name and press F5 YES key F5 If the file is protected the file can not be erased Erase the file after canceling the protection Ope...

Страница 91: ...The Baud rate selections are 600 1200 2400 4800 9600 and 19200 Operating procedure Operation Display 1 Press F1 Protocol key to set protocol F1 2 To select the Baud rate use the arrow key s F3 F6 to h...

Страница 92: ...mand the PC to send the file at this time Press the F1 START key to start transferring File name amount of received data Byte Capacity of the file Byte and percentage of proceeding will be displayed W...

Страница 93: ...sferring 3 File name amount of sent data Byte Capacity of the file Byte and percentage of proceeding will be displayed Each file selected will be transferred in order When the transfer is complete the...

Страница 94: ...ent of Compensation Systematic Error of Instrument and Section 8 5 Showing Constant List and Switch ON OFF Compensation Systematic Error of Instrument 5 V 0 Zenith Level Select the vertical angle read...

Страница 95: ...it for the atmospheric correction 16 Pres Unit mmHg inHg hPa Select the air pressure unit for the atmospheric correction 17 R L Lock OFF ON Prohibit switching angle right or left by soft key in angle...

Страница 96: ...CRLF Select the option OFF or ON for carriage return and line feed when collecting measurement data with a computer 6 REC A B A B Select the option to record the data REC A The measurement is started...

Страница 97: ...and select hPa F3 7 Press F1 SET key F1 8 Press F5 YES key 1 The display returns main menu F5 1 To cancel the setting press F6 NO key Parameters F1 Measurement F2 Communication F3 Password Parameters...

Страница 98: ...as Section 7 2 1 Parameters for Measurement and Display refer to the section 7 2 1 1 1 Press the F2 I GTS key to reset to the factory default settings Topcon total station current fixed protocol Fact...

Страница 99: ...arameters option menu To access the password option press F3 Password F6 2 The password option screen will appear When accessing the password option for the first time the OFF indicator will appear on...

Страница 100: ...ccess the parameters option menu Press F3 Password to access the password option F6 F3 2 Type in your password and press ENT 3 The password option screen will appear The indicator on the upper left si...

Страница 101: ...ange to change the current password Type password ENT F2 4 The password input screen will appear At the blinking cursor type in the new password and press the ENT key IMPORTANT DON T FORGET YOUR PASSW...

Страница 102: ...differences in temperature greatly affect the length measured If a difference of 5mm or over is the result from the comparative measurement the following procedure as shown below could be used to chan...

Страница 103: ...audio 4 Sight through the eyepiece and focus the blinking red light spot by turning the focusing knob in the direction of infinity clockwise If displacement of the reticle cross hairs is within one f...

Страница 104: ...ten by revolving in the direction of tightening tension 4 The attachment screws must also be tightened sufficiently upon completion of adjustments 5 Always repeat checking operations after adjustments...

Страница 105: ...by this method 2 Correct the remaining amount of the bubble displacement with the leveling screws 3 Rotate the instrument 180 or 200g around the vertical axis once more and check bubble movement If th...

Страница 106: ...er if the point appears to be displaced from the vertical cross hair as the telescope is swung vertically then proceed with the following adjustment Adjustment 1 Unscrew the cross hair adjustment sect...

Страница 107: ...ore and tighten the horizontal motion clamp 7 Loosen the vertical motion clamp only and rotate the telescope 180 or 200g around the horizontal axis once more and fix a sight on Point C which should co...

Страница 108: ...0 or 200g around the vertical axis once more and check the center mark If it is coincided to the point then further adjustment is not required Otherwise repeat the adjustment Note 1 First loosen the c...

Страница 109: ...eveling screws and coincide the center of the laser and the point 4 Rotate the instrument 180 or 200g around the vertical axis once more and check the point and the laser If the laser is coincided to...

Страница 110: ...so far Operating procedure Operation Display 1 Level the instrument properly with the plate level 2 Press F5 key from the main menu F5 3 Press F1 key F1 4 Collimate target A around 0 in horizontal wi...

Страница 111: ...lay returns to main menu F6 1 It is able to get the average value from 1 to 10 measurements To get the average repeat the procedures in steps 4 5 or 9 10 The measured times is counted in the second li...

Страница 112: ...procedure Operation Display 1 Press F5 key from the main menu F5 2 Press F2 key Correction values are displayed F2 3 Press F6 OFF key F6 4 Press F1 EXIT key The display returns to main menu F1 Adjust...

Страница 113: ...s ENT key Example 13 20 50 The display returns to main menu 0 1 2 9 9 5 ENT Enables you to change the order of date refer to Chapter 7 PARAMETERS SETTING MODE Adjustment F1 V0 Axis Measurement F2 V0 A...

Страница 114: ...2 Press F4 Instrument constant key GTS 600AF series needs to get to next page by pressing the F6 key F4 3 Press F5 YES key F5 4 Input value and press ENT key Input value ENT 5 Press F5 OK key The disp...

Страница 115: ...ng procedure Operation Display 1 Press the F5 key from the main menu F5 2 Press the F6 key to get to next page on the display F6 3 Press the F1 key The beam will be emitted F1 4 To return to the main...

Страница 116: ...ode Setting example The prism constant value 14mm Operating procedure Operation Display 1 Press STAR key 2 Press F6 key to get the function as in page 2 F6 3 Press F3 key Current setting value is disp...

Страница 117: ...ospheric correction is not set Example In case Temperature 20 C Air pressure 847hPa l 1000 m 50 10 6 50 ppm L 1000 1 50 10 6 1000 050 m 10 2 Setting of Atmospheric Correction Value How to Set Temperat...

Страница 118: ...calculated from the input temperature and pressure values exceeds the range 999 9ppm the operating procedure returns to step 3 automatically Input values again Operating procedure Operation Display 1...

Страница 119: ...ly with the atmospheric correction chart Find the measured temperature in horizontal and pressure in vertical on the chart Read the value from the diagonal line which represents the required atmospher...

Страница 120: ...10 SETTING ATMOSPHERIC CORRECTION 10 4...

Страница 121: ...10 SETTING ATMOSPHERIC CORRECTION 10 5...

Страница 122: ...Horizontal distance D AC or BE Vertical distance Z BC or EA D L cos 2 sin Z L sin cos L cos 2R Earth curvature correcting item K Lcos 2R Atmospheric refraction correcting item K 0 14 or 0 2 Coefficie...

Страница 123: ...le battery can be used repeatedly by charging If charging is repeated by the state that capacity of the battery still left operating time of the battery may shorten In this case the voltage of the bat...

Страница 124: ...of 10 C to 40 C 50 F to 104 F 2 If charging is done at high temperature charging time of the battery may take longer 3 Exceeding the specified charging time may shorten the life of the battery and sh...

Страница 125: ...ment by the carrying handle with one hand and carefully lower it on top of the tribrach while at the same time coinciding the alignment piece with the tribrach alignment groove on the instrument and t...

Страница 126: ...time approx 15 hour 20 C to charge BT 3L Operation temperature range 10 C to 40 C 50 to 104 F External dimensions 142 L 96 W 64 H mm Weight 1 0kg Large capacity battery pack BT 3L Output voltage DC 8...

Страница 127: ...rnal dimensions 116 L 60 W 50 H mm Weight 0 3kg Diagonal eyepiece Model 10 Observation in an easy posture will be provided up to the zenith position Trough compass Model 6 Shock proof construction No...

Страница 128: ...kg Gadget case Model 1 A case to store and carry accessories External dimensions 300 L 145 W 220 H mm Weight 1 4kg Prism unit case Model 5 1 prisms unit or fixed 3 prisms unit can be stored in this ca...

Страница 129: ...14 SPECIAL ACCESSORIES 14 4 Aluminum extension leg tripod Type E Flat head 5 8 11 threads with adjustable legs Wide frame extension leg tripod Type E Wood Flat head 5 8 11 threads with adjustable legs...

Страница 130: ...3Q BC 10B BC 10C Approx 15h BC 5 BC 9 BT 3L BC 6 Normal BC 10B for AC 120V use BC 10C for AC 230V use Normal for AC 100V 120V 220V 240V use Quick for DC 13 8 to 16V use Quick for AC 100V 120V 220V 24...

Страница 131: ...height of prism set change the position of 4 fixing screws Target pole 2 not used with 9 prisms Prism 2 Tilting prism holder 2 with target plate 2 Pole adaptor A Pole adaptor F2 Tribrach Optical plum...

Страница 132: ...ying case and taking out the instrument place the case horizontally then open the case 7 When returning the instrument to its case be sure to match the white positioning marks provided with the case a...

Страница 133: ...error Displayed when the telescope rotated too fast or any abnormality occurs in angle measuring system The instrument will return to previous mode automatically E31 Displayed when the unit of the ang...

Страница 134: ...ration procedure is correct E90 s Abnormality in internal memory system Repair is required E400 s Abnormality in auto focusing system Only for GTS 600AF series Retry the auto focus by pressing the aut...

Страница 135: ...ent mode 1mm 0 005ft Tracking measurement mode 10mm 0 02ft Measurement Display 11 digits max display 9999999 9999m Measurement Time Fine measurement mode 1mm 1 2sec Initial 4 sec 0 2mm 2 8sec Initial...

Страница 136: ...ection unit 1 0 1mgon Computer unit OS MS DOS Ver 3 22 Internal Memory System memory FEEPROM 512KB Main memory RAM 640 KB Data memory RAM 320 KB Program memory FEEPROM 512KB Application program memory...

Страница 137: ...ratio of 1 distance measurement 3 angle measurement GTS 600AF series The automatic focusing is used once in every 30 seconds Including distance measurement 5hours Angle measurement only 9hours Normal...

Страница 138: ...rget is 10 above the horizon and rotated 90 in azimuth from the direction of the vertical axis error Hzerr 30 sin tan10 Hzerr 30 1 0 176326 5 29 From the above example it can be seen that horizontal a...

Страница 139: ...ion it is necessary to keep the compensators in proper adjustment The compensators must agree with the actual level condition of the instrument Through various environmental stresses the agreement bet...

Страница 140: ...d 3 Storage See Fig 3 for how storing period at different temperature levels is related to the remaining capacity The battery will lose its capacity as storage temperature rises and the storage period...

Страница 141: ...g Discharge 0 2 E V 1 0V Temperature at discharge 20 0 20 40 20 40 60 80 100 120 0 Ambient temperature Discharge capacity 0 2C discharge Fig 2 Discharge 1C discharge 60 20 Charging 0 1C 16hours Chargi...

Страница 142: ...r provided in the documentation for the embedded system Should you have any questions concerning this Agreement or if you desire to contact Embedded System Manufacturer for any other reason please ref...

Страница 143: ......

Страница 144: ...Dongcheng District Beijing 100027 China Phone 10 6501 4191 2 Fax 10 6501 4190 TOPCON CORPORATION BEIRUT OFFICE P O BOX 70 1002 Antelias BEIRUT LEBANON Phone 961 4 523525 961 4 523526 Fax 961 4 521119...

Отзывы: