Example:
:TEST:DATAI
?
If the present monitor current in an ACW test is 1.00,
1.00 is returned.
:
TEST:DATAR
?
Inquires about the monitor resistance during the IR test.
--Syntax
:
:TEST:DATAR
?
--
Unit: M
Ω
--
Example:
:TEST:DATAR
?
If the present monitor resistance in an IR test is 1.00,
1.00 is returned.
:
TEST:FETCH
?
Inquires the test results of per step.
--Syntax
:
:TEST:FETCH
?
--
Response message:
data1, data 2,…, datan
--
Unit: mA/M
Ω
--
Example:
TEST:FETCH
?
If the result of step1 is 1.00mA, the step2’s is 3.00mA,
1.00,3.00 is returned.
:
TEST:FETCH2
?
Inquires the test voltage, current (for ACW or DCW test)or resistance(for IR test) ,status of the
current step.
--Syntax
:
:TEST:FETCH2
?
--
Response message:
Status value, voltage value, current or resistance value
----Program data
< status >:
0
(
READY
)
1
(
TEST
)
2
(
PASS
)
3
(
FAIL
)