26
UPPER:
1.000 mA
FAIL:
0.5 S
OFST
REAL:
1.000 mA
LOW:
0.001 mA
RMT
TIME:
1.0 S
FREQ:
50 Hz
ERR
ARC:
5.00 mA
INS
DEL
NEW
F1 F2 F3 F4 F5
Figure 4.2.1
AC setup interface
Instruction
STEP: 01/01
Test procedure
:
current setting No. / total items.
Key Function
Instruction
F1 INS Within the current test programme
(
PROG
),
it adds a new test
item.
F2 DEL Within the current test programme
(
PROG
),
delete the current test
item.
F3 NEW
Create a blank test programme
(
PROG
)(
including a new test
item
),
which is used to write a new test programme.
ENTRY
Switch the test item within
(
PROG
)
and select the item needing
changing.
START
Directly setup the measurement
AC
Current operating mode of test procedure is AC withstanding voltage.
Cursor in this position can switch to other working items by F1~F3, for instance
DC
、
IR
、
OS
.
SCAN: 1X2X3X4X5X6X7X8X Connect multi-channel output and test terminal.
SCAN channel
:
as for TH9201S with multi-channel
,
other models are corresponded with
output terminal of TH9121multi-channelstester. 8 output channels is connected with test output
terminal
,
there are open, high voltage output and ground low terminal.
Other data The test condition of current
(
AC
)
item
(
in 4.3 AC withstanding voltage
setup
)
.
4.2.2
(
TEST
):(
Take AC for example
)
STEP: 01/01
AC
SCAN:
1X2X3X4X5X6X7X8X
VOLT: 1.000
kV
LOCK
UPPER: 1.000 mA
1 0 0 0 VAC
OFST
REAL: 1.000
mA
RMT
TIME: 1.0 S
1.0 0 0 mA
ERR
ARC: 1.00
mA
0.5 S
READY
Figure 4.2.2
AC test
Note
:
Press
T
to enter set interface
The high voltage can only be started on this interface to test high voltage. Other test
conditions have to been set on setup interface. The test procedure is as below
: