Unit: s
--
Example:
To set the WAIT time for step 1 to 1,
:SOUR:SAFE:STEP 1:DC:TIME:DWEL 1.
--
Example:
:SOUR:SAFE:STEP 1:DC:TIME:DWEL?
If the current WAIT time for step 1 is 1,
1 is returned.
:SOUR:SAFE:STEP:DC:CLOW
Sets the CHECK LOW’s status for DCW test.
Also inquires about the CHECK LOW’s status.
--Syntax
:
Command message:
SOUR:SAFE:STEP <sn>:DC:CLOW <ON/OFF> or <1/0>
Query message:
SOUR:SAFE:STEP <sn>:DC: CLOW ?
--Program data
<sn>:
Data format: Integer
Set value: 1~49
Resolution: 1
Unit:
--Program data
<set value>:
Data format: Character
Set value: OFF(0), ON(1)
Resolution:
Unit:
--
Example:
To set the CHECK LOW’s status for step 1 to ON,
:SOUR:SAFE:STEP :DC:CLOW ON.
--
Example:
:SOUR:SAFE:STEP :DC:CLOW?
If the current CHECK LOW’s status for step 1 is ON,
ON is returned.
:SOUR:SAFE:STEP:DC:CHAN
Sets HIGH/LOW/OPEN for the scanner channel for DCW test.
Also inquires about the set value for the scanner channel.
--Syntax
: