![Tonghui Electronics TH9201 Series Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/tonghui-electronics/th9201-series/th9201-series_operation-manual_1126463047.webp)
◇
47
:SOUR:SAFE:STEP:AC:LIM:REAL 0.001.
--
Example:
:SOUR:SAFE:STEP:AC:LIM:REAL?
If the current REAL current for step 1 is 0.001,
0.001 is returned.
:SOURce:SAFEty:STEP:AC:TIME:RAMP
Sets the RISE time for ACW test.
Also inquires about the RISE time.
--Syntax
:
Command message:
SOUR:SAFE:STEP <sn>:AC:TIME:RAMP <time>
Query message:
SOUR:SAFE:STEP <sn>:AC:TIME:RAMP?
--Program data
<sn>:
Data format: Integer
Set value: 1~49
Resolution: 1
Unit:
--Program data
<time>:
Data format: float
Set value: 0~999.9 (0 is OFF)
Resolution: 0.1
Unit: s
--
Example:
To set the RISE time for step 1 to 1,
:SOUR:SAFE:STEP 1:AC:TIME:RAMP 1.
--
Example:
:SOUR:SAFE:STEP 1:AC:TIME:RAMP?
If the current RISE time for step 1 is 1,
1 is returned
.
:SOURce:SAFEty:STEP:AC:TIME:FALL
Sets the FALL time for ACW test.
Also inquires about the FALL time.
--Syntax
:
Command message:
SOUR:SAFE:STEP <sn>:AC:TIME:FALL <time>
Query message: