![Tonghui Electronics TH9201 Series Скачать руководство пользователя страница 63](http://html1.mh-extra.com/html/tonghui-electronics/th9201-series/th9201-series_operation-manual_1126463063.webp)
◇
63
Resolution: 1
Unit:
--Program data
<HIGH/LOW/OPEN>:
Data format: Character
Set value: HIGH/LOW/OPEN
Resolution:
Unit:
--
Example:
To set the scanner channel 1 for step 1 to HIGH,
:
SOUR:SAFE:STEP 1:IR:CHAN 1:HIGH.
--
Example:
SOUR:SAFE:STEP 1:IR:CHAN 1?
If the scanner channel 1 for step 1 is HIGH,
HIGH is returned
.
5.
Commands for OS Settings
:
SOUR:SAFE:STEP:OSC:OPEN
Sets the OPEN rate for OS test.
Also inquires the OPEN rate.
--Syntax
:
Command message: :
SOUR:SAFE:STEP <sn>:OSC:OPEN <set value>
Query message:
:SOUR:SAFE:STEP <sn>:OSC:OPEN?
--Program data
<sn>:
Data format: Integer
Set value: 1~49
Resolution: 1
Unit:
--Program data
<set value>:
Data format: float
Set value: 0.1~1.0
Resolution: 0.1
Unit:
--
Example:
To set the OPEN rate for step 1 to 50%,