8
Instructions for Use
TNI
soft
Flow
• Do not let children play with the hoses or cables to prevent injuries such as strangulation or swallo-
wing of small parts.
• Do not reach into the housing directly after use since inner parts such as the heating plate and the
bottom of the humidifier might be hot.
• Do not cover the device nor the applicator during use.
• Position the device where it cannot fall into water.
• Disconnect the power supply and discontinue use if water enters the housing or escapes from the
humidification chamber.
• Disconnect the power supply and discontinue use if the device has been dropped or damaged.
• If the power cord or plug is damaged, disconnect the power supply and discontinue use.
• Do not use damaged applicators.
• Do not connect unauthorized components to the power socket.
• Ensure that the characteristics of the local power supply correspond to the requirements of the TNI
softFlow system. See the device’s name plate and chapter Product specifications.
• Do not supply any gases other than oxygen via the lateral oxygen inlet port.
• Do not use the TNI softFlow system in MRI environments, near HF surgical equipment or in other
environments where the intensity of the EM disturbances is high.
• Do not use the TNI softFlow at high altitudes as the therapy quality might be affected negatively
due to low ambient pressure.
• If existing, do not remove the protective caps from the accesses below the carrying handle! The
accesses are intended exclusively for maintenance purposes. Before commissioning, ensure that the
connections are tightly closed by the protective caps.
Contraindications
• Do not use the TNI softFlow applicators if you are allergic to silicon.
• The TNI softFlow system is not intended as a life-supporting measure.
• The TNI softFlow system may not be used for invasive ventilation.
• The nasal application of high flow by means of TNI softFlow must not be used if the patient’s upper
respiratory tract is completely obstructed.
• The nasal application of high flow by means of TNI softFlow must not be used if the patient’s upper
respiratory tract has been bypassed using a bypass.
• TNI softFlow system must not be used in patients who have a history of anamnestic seizures or
restless sleep.
• If the patient‘s upper respiratory tract is bypassed, only use the tracheal interface applicator with an
open tracheal interface.
• Do not use the applicator headgear if you have significant pressure marks from it or in the event of
material incompatibility of the stretch band or applicator headgear.