![TNI softFlow Скачать руководство пользователя страница 39](http://html.mh-extra.com/html/tni/softflow/softflow_instructions-for-use-manual_1122046039.webp)
39
Instructions for Use
TNI
soft
Flow
Instructions for Use
TNI
soft
Flow
5.2 System information
Clinic mode
The TNI softFlow is delivered in clinic mode; the ‘clinic menu’ is activated. The qualified health professional
is enabled to set all therapy parameters and system configurations and to access the menu item ‘system
information’ which provides additional information such as firmware versions and serial number.
WARNING
• If a TNI softFlow is used at home, the health care personnel must deactivate the clinic menu
after setting the therapy parameters so that the patient cannot change the settings during
operation. This procedure requires a PIN code. Please contact your TNI medical AG represen-
tative if you intend to deactivate the clinic menu.
NOTE
•
The “Service menu” can be accessed by a TNI medical AG technical staff or representative only.
Homecare mode
If the TNI softFlow is used at home, its configuration needs to be changed to homecare mode to prevent
accidental access to therapy parameters. To unlock such a device again (once or permanently) and to access
all parameters please contact your TNI medical AG representative.
5.3 Ambient conditions
See chapter Product specifications: Environmental conditions
Ambient temperature:
10 - 30°C
Ambient humidity:
15 - 93% RH
Altitude:
0 – 3000 m MSL
WARNING
• Operating the device at high altitudes may have negative impact on therapy quality.
Storage and transport conditions
See chapter Product specifications: Environmental conditions concerning storage and transport.
The device should be stored and transported at temperatures between -29 to +70°C, ≤93% RH, 700 –
1060 hPa. The device may be transported only in upright position and if completely dry.
List of compatible gases
Room air enters the device through the air slot and the dust filter at the rear side.
Oxygen is provided by an external oxygen source.
5.4 Data storage
All data on operation and dysfunction of the TNI softFlow system are recorded during therapy hours and
can be accessed by TNI medical AG technical staff. These data are stored in the internal memory. The inter
-
nal memory has the capacity to store all data collected during the previous 12 months. Additionally, an SD
card can be used to save data independently of the internal memory.