3.2 Therapy parameters
In operation mode, the display shows the current output humidity (dew point temperature in °C DP), fl ow
rate (in l/min), oxygen fl ow rate (in l/min) and FiO
2
(in %).
Numbers in the bottom line show the programmed nominal values. Arrows in front of the output values
indicate that the nominal values are not reached yet and the device is currently up- or down-regulating the
respective parameter.
CURRENT VALUE
NOMINAL VALUES
CURRENT VALUE
3.2.1 Flow rate
• Select the parameter ”Flow” in the user menu.
• Adjust the flow rate in 0.5 l/min steps to the value, that is suitable to the particular applicator type,
and confirm the selected nominal value.
• The newly set nominal value is shown in the footer at the bottom of the display.
User menu:
Setting:
Nominal value in the footer:
NOTE
•
Set the fl ow rate before attaching the applicator to the patient to prevent discomfort.
WARNING
•
Flow rates must be set by qualifi ed health professionals only. For this reason this menu item is
hidden in homecare mode.
3.2.2 Humidity
• Select the parameter ”Humidity” in the user menu.
• Increase the dew point temperature (in 1°C DP steps, within the range from 30-37°C DP) to increase
the humidity or vice versa by pressing the arrow keys. Confirm the new nominal value.
• The newly set nominal value is shown at the bottom of the display.
User menu:
Setting:
Nominal value in the footer: