![TNI softFlow Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/tni/softflow/softflow_instructions-for-use-manual_1122046029.webp)
29
Instructions for Use
TNI
soft
Flow
Instructions for Use
TNI
soft
Flow
3.2.3 Oxygen
If required, oxygen can be additionally mixed into the air fl ow by connecting an external oxygen source to
the TNI softFlow device (see chapter Oxygen supply).
• Switch on the TNI softFlow device first.
• Wait until the target flow is reached.
• Start the oxygen supply by opening the valve of the external oxygen source.
• The oxygen flow rate is displayed in l/min and the resulting oxygen concentration of the air flow is
shown as FiO
2
value in %.
• Adjust the oxygen supply by adjusting the opening of the valve of the external oxygen source.
• Stop the oxygen supply by closing the valve of the external oxygen source.
Display of the oxygen fl ow rate and the FiO
2
NOTE
• The admixture of oxygen is limited to 20 liter per minute using an applicator tracheal interface
• When starting or restarting the device make sure the oxygen supply is closed. Slowly open the
oxygen supply after the device has reached set air fl ow completely.
• If the FiO
2
exceeds 95% a value of 99% will be displayed.
WARNING
•
Smoking and open fi re are strictly prohibited when using supplemental oxygen due to the risk of explosion.
• Do not place a connected applicator on the TNI softFlow device or any other electrically
driven device when the device is running.
• Keep a min. distance of 1 m from other electrical devices when using oxygen.
• Alternatively, the nominal value can be changed directly in the operation mode by pressing the arrow keys.
WARNING
• During tracheal application the humidity is automatically set to 37°C DP and cannot be
changed as long as the tracheal applicator is in use.
NOTE
•
For optimal humidifi cation of the patient’s mucosa, humidity of 34-37°C DP during therapy is
recommended.
•
If the patient feels dryness in the nose, check if the humidifi cation chamber contains enough
water and/or increase the humidity value.
• The system requires a setup-time of about 10 min to adjust a newly set nominal value of humidity.
•
If water condenses excessively in the applicator / heating tube, the chosen humidity value
might be too high for the present ambient conditions. Reduce the dew point value.
• When operated in junior mode condensation in the cannula may occur in certain ambient
conditions at fl ow rates less than 5 l/min. To minimize condensation, it is recommended not
to set the humidity higher than 34°TP, if using fl ow rates less than 5 l/min.
•
Humidifi cation system output is >12 mg / l for any setting during nasal application.
•
Humidifi cation system output is >33 mg / l for any setting during tracheal application.