Perform measurement
4.4
Connect measurement Leads
Connect the Patient Ground lead to the GND input of the amplifier and to the
subject. Use the TMSi Patient Ground wristband for optimal contact. Wet the band
and place it around the wrist.
TMSi recommends using the wet wristband to optimize
measurement setup. This will improve the signal quality. Please
read the Application Note on the reference amplifier for more
background. (
Connect all measurement leads and Patient Ground. Please refer to instructions for
use of the accessories and sensors for more information.
TMSi Polybench: Quick Recording Application
The Quick Recording Application is a TMSi Polybench measurement configuration
supplied with your product or downloadable on
In case you are not using TMSi Polybench, but a different
software package in combination with our products, please refer
to the User Manual of that software package.
Using impedance measurement
4.5
The impedance measurement of the Refa is controlled by the commands from the
application software. Refer to instructions for use of the application software to use
impedance mode on the Refa.
The device requires the following to be able to measure the impedance correctly:
-
At least one odd and one even channel above channel number 4 must be
connected
-
Connected Patient Ground electrode.
-
The system must be set to impedance mode.
Until all these requirements are fulfilled, impedance LEDs will blink to indicate that
impedances cannot be determined properly.
Please follow the instructions for use for the TMSi Polybench Quick Recording
Application. For other application software, refer to the user manual of that
software.