Rev. 6/06/18
•
When moving on to the next weight REMOVE ALL WEIGHT AND RE-ZERO before adding
more weight.
•
Example: for this screen after you press the check mark it will ask for 20 grams. Remove the 10
grams and re-zero, then add the 20 grams.
•
Repeat this process for all of the instructed weights.
•
After the calibration is complete, the Final Review screen will appear.
•
Cycle through the values, review and record them from the table, since they will be put on the
calibration record.
•
If, for whatever reason, one value is wrong tap that value and you can re-calibrate that specific
weight.
•
Also record the information in the red box. This information will also go on the calibration record.
•
The
se values can be accessed from the load cell menu’s “View” button.
•
When complete press ACCEPT.