![TMI Products TQC Sheen 32-76e Series Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/tmi-products/tqc-sheen-32-76e-series/tqc-sheen-32-76e-series_manual_1121874036.webp)
32-76-00-0002
36
Rev. 6/06/18
•
You will need a calibrated weight set (shown below) for the next part.
•
The machine will inform you what weight you will need in the recommended calibration weight
series. If you are calibrating with a custom weight set you can c
hange the “Target Force” by
pressing the box on the screen that contains the desired value. A prompt will open for you to enter
a new gram value.
•
Before hanging the weight, press the zero key and allow the unit to get as close to zero as possible.
(this may take more than one press)
•
Hang the weight on the hook. Let the current force value settle.
•
Adjust the current force value with the up and down buttons so it is as close as possible to the
target force.
•
When the desired value is reached press the
GREEN
check to accept the calibration point (Current
Force is acceptable) and move on to the next weight. Touch the
RED
X to skip the current and
any remaining weights and proceed to the final review screen.