10. Optional Settings and Adjustments
072-65284M-02 (SonicMan SN1) 24 Feb 2005
22
10.9 Setting the Automatic Power OFF Time Delay
(Factory default is “10 min”)
This feature automatically turns the
SonicMan
power OFF, after the prescribed length of
time has elapsed in which the
SonicMan
has not been operated.
1) Press
the
[
×
] and [
Ø
] buttons to raise or lower the power
OFF time delay from 0 to 60 minutes.
2) If the time delay is set to 0 minutes, the automatic power
OFF feature is disabled.
3) Press [ENT] to move the display to the next setting item.
4) Press [Func.] to return to detection mode.
10.10 Adjusting the Earphone Output Volume
(Factory default is “Maximum”)
This feature allows the adjustment of the earphone output volume. As there is no level
indicator, perform the adjustment while monitoring the sound from the earphones.
1) Pressing either the [
×
] or the [
Ø
] button one time adjusts
the level one step. If the button is pressed and held down,
the level is continuously adjusted until the button is
released. To go from the minimum to the maximum, the
button must be held down continuously for approximately 7
seconds.
2) Press [ENT] to move the display to the next setting item.
3) Press [Func.] to return to detection mode.
10.11 Adjusting the Display Contrast
(Factory default is “Maximum”)
This feature allows the screen contrast to be adjusted. As there is no level indicator,
perform the adjustment while observing the screen.
1) Pressing either the [
×
] or the [
Ø
] button one time adjusts
the level one step. If the button is pressed and held down,
the level is continuously adjusted until the button is
released. To go from the minimum to the maximum, the
button must be held down continuously for approximately 7
seconds.
2) Press [ENT] to move the display to the next setting item.
3) Press [Func.] to return to detection mode.
10.12 Unit Setting
(Factory default is “SI”)
This feature is used to set the units for calculating detected
value, switching between “SI” (m for length and kg for weight)
and “Imperial” (ft for length and lb for weight).
1) Pressing either the [
×
] or the [
Ø
] button one time switches
between “SI” and “Imperial”.
2) Press [ENT] to return to the battery strength display.
3) Press [Func.] to return to detection mode.
Auto OFF
*
*
*
*
10 min
ENT
Volume
*
**
*
:UP :DOWN
ENT
Contrast
*
*
*
*
:UP :DOWN
ENT
Unit
SI
ENT