8. Estimating Leak Level, Saving and Recalling Data
072-65284M-02 (SonicMan SN1) 24 Feb 2005
16
8 Estimating Leak Level, Saving and Recalling Data
8.1
Hold the Inspection Data
1) Press the [Memory] button while detecting a leak.
2) If the automatic leak level calculation function is set to “Auto” (see section 10.4
“Setting Leak Level Calculation”), the display changes to the screen for input of the
estimated leak level calculation conditions.
3) If the automatic leak level calculation function is set to “Manual” (see section 10.4
“Setting Leak Level Calculation”), the display changes to the record number selection
screen without calculating the leak level.
8.2
Calculate Leak Level
Leak level can be calculated after saving the
SonicMan
reading (see section 8.5
“Recalling Saved Data”).
1) Enter the distance from the leak to
SonicMan
’s lead
edge.
(a) Pressing either the [
×
] or the [
Ø
] button 1 time
increases or decreases the distance by 0.1 m,
respectively. Pressing and holding increases or
decreases the distance in 1.0 m increments.
The distance input range is from 0.1 m to 25.5 m.
(b) Press the [ENT] button to confirm the data and
then move to the next data item.
(c) When the nozzle is attached, the distance to the
front the nozzle is 0.1 m.
2) Select Fluid Type. Select either “Steam” or “Gas”.
(a) Scroll through the choices for fluid type by
pressing the [
×
] and [
Ø
] buttons.
(b) Select the type of fluid that is leaking, then press
the [ENT] button.
(c) The result from the leakage amount calculation
is displayed, then the display changes to the
record number confirmation screen.
(d) For revising Calculation Conditions:
Pressing the [Func.] button again allows for re-
input of the calculation conditions as many times
as desired.
3) Canceling the Input of Calculation Conditions before Input is Complete
Pressing the [Func.] button causes the display to skip the conditions input screen and
to jump to the record number confirmation screen.
Data Hold
Memory
Calculate Leak Level
(When Cal. function is “Auto”)
ENT
Select Record
Number
Save Data
Memory
(When Cal. function is “Manual”)
Input Distance
No.001 *
No.001 *
No.001 *
No.001 *
Distance 0.1m
Distance 0.1m
Distance 0.1m
Distance 0.1m
No.001
*
*
*
*
Distance 0.1m
Select Fluid Type
No.001 *
No.001 *
No.001 *
No.001 *
Fluid
Fluid
Fluid
Fluid
エアー
エアー
エアー
エアー
No.001
*
**
*
Fluid Gas
ENT
Display Calculation Result
No.001 *
No.001 *
No.001 *
No.001 *
L:
L:
L:
L:
---
---
---
---
60dB S 50
60dB S 50
60dB S 50
60dB S 50
No.001
*
*
*
*
L
:
::
:
1
60 S 50
ENT