EF200W-C
15
Low flow cut off
The switch points for low flow cut off are preset and can be configured.
Galvanic isolation
All inputs and outputs are galvanically isolated from one another.
Protocol-specific data
HART
Manufacturer ID
0x11
Device type ID
0x0038
HART protocol revision
7
Device description files
(DTM, DD )
Consult TLV for more information.
HART load
• Min. 250 Ω
• Max. 500 Ω
Power supply
Terminal assignment
Transmitter
Connection versions
3
4
1
2
C
+
‒
+
‒
A
B
A0033475
Maximum number of terminals
Terminals 1 to 4:
Without integrated overvoltage protection
A
B
C
Output 1 (passive): supply voltage and signal transmission
Output 2 (passive): supply voltage and signal transmission
Ground terminal for cable shield
Terminal numbers
Output 1
Output 2
1 (+)
2 (-)
3 (+)
4 (-)
4-20 mA HART (passive)
Pulse/frequency/switch
output (passive)
Output 1 must always be used; output 2 is optional.