EF200W-C
50
Operating elements
• Operation with 3 push buttons with open housing: , ,
Additional functionality
• Data backup function
The device configuration can be saved in the display module.
• Data comparison function
The device configuration saved in the display module can be compared to the current device
configuration.
• Data transfer function
The transmitter configuration can be transmitted to another device using the display module.
Remote operation
Via HART protocol
This communication interface is available in device versions with a HART output.
1
10
2
3
4
5
7
6
9
8
A0028746
Fig. 21
Options for remote operation via HART protocol (passive)
1
Control system (e.g. PLC)
2
Transmitter power supply unit, e.g. RN221N (with communication resistor)
3
Connection for Commubox FXA195 and Field Communicator 475
4
Field Communicator 475
5
Computer with web browser (e.g. Internet Explorer) for accessing computers with operating tool (e.g.
FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) with COM DTM "CDI Communication TCP/IP"
6
Commubox FXA195 (USB)
7
Field Xpert SFX350 or SFX370
8
VIATOR Bluetooth modem with connecting cable
9
Transmitter