LabTower EDI
Ultrapure water system
10
6. Technical specifications
Demands made on the raw water
Source
Potable tap water softened or hardness stabilized.
Blocking index (SDI)
< 3, with higher values, an upstream pretreatment
(article no. 09.4000) is to be installed.
Conductivity
<1500 µS/cm
Prefiltration
5µm + activated hardness stabilization
Free chlorine
< 0.1 mg/Litre
Manganese content
< 0.05 mg/Litre
Iron content
< 0.05 mg/Litre
Colloid index
< 3
ph-Range
4 - 11
Temperature
2 - 35 °C
Pressure
2 - 6 bar
Product water quality ASTM I (dispensing valve)
Standard
UV (option)
Conductivity
µS/cm
0.055
0.055
Resistance
M
!
xcm bei 25°C
18.2
18.2
TOC
ppb
5 - 10
1 - 5
Bacteria
KBE/ml
< 1
< 1
Particles
> 0,2 µm
< 1/ml
< 1/ml
Flow rate
L/min
1.5
1.5
Typical
volume taken
10 L/day
Product water quality ASTM II (tank quality)
LabTower EDI 15
LabTower EDI 30
LabTower EDI 30
Retention quota, salts
Ø 98 %
Ø 98 %
Ø 98 %
Retention quota, bacteria and
particles
99 %
99 %
99 %
Conductivity
0.067 – 0.1 µS/cm 0.067 – 0.1 µS/cm 0.067 – 0.1 µS/cm
Resistance
M
!
xcm bei 25°C
15 – 10
15 - 10
15 - 10
Performance
15 L/h
30 L/h
45 L/h
Dimensions
Height
1500 mm
Width
450 mm
Depth
580 mm
Weight, empty
58 kg
Operational weight (with full tank)
158 kg