1. Place the Ultrapure water system on the table of the storage tank. Connect the hoses
positioned at the table by putting the connectors in with the appropriate connections (19-23) at
the equipment.
2. Remove the covering hood (6) of the storage tank.
3. Turn the filter housing of the pre-treatment (13) loosely. Insert the pre-treatment unit con-
tained when assembling (hardness stabilization, pre-filter and connector) into the filter
housing and screw this again firmly (see chapter „Change of the pre-treatment “).
4. In the case of use of the optional sterile or fine filter you turn the filter housing (14) loosely.
Insert the filter into the filter housing and bolt this on again.
5. Connect the Ø8 – ¾“ female thread feedwater hose supplied to the feedwater inlet (7) of
the storage tank and to a water tap that can be turned off. The connecting hose is marked
“Feedwater“.
6. Connect the Ø8 concentrate hose supplied to the concentrate outlet (8) of the storage
tank and to the drain provided on-site. The concentrate hose is marked „Concentrate“.
Caution! The concentrate must be able to flow under free gravity fall to the drain!
7. Plug the 4-pin plug of the level control system (storage tank), marked „Niveau“, in the 4-pin
socket (18) of the Ultrapure water system and screw the milled nut of the plug tight.
Connect the plug (power supply EDI, 230 V), marked “EDI 230V”, to coupling (17) of the
Ultrapure water system. Fix the plug by firmly pressing the clamp.
8. Screw the optional ultra filtration capsule (4) into the withdrawal tap G1/4 “.
9. Now fit the hose of the Ø8 mm sterile overflow to the overflow on the back of the tank and
connect it to the drain.
10. Ensure that residual water drainage (10) is turned closed.
11. Check the feedwater pressure; it must be within the permissible pressure range
(see Technical specifications).
Before you now switch the system on, read through the rinsing out
procedure for reverse osmosis membranes packed with preserving
agent in the „Rinsing the membrane“!
12. Plug the line plug in and switch the Ultrapure water system on with the on/off key on the
operating unit (5).
13. Check that all connections are leak-tight.
14. Replace the covering hood on the storage tank.
After a short rinsing phase, your system begins to produce pure water into the storage tank.