8. The installation area
The following criteria must be taken into consideration when choosing the installation area.
#
Minimum temperature in the installation area: + 2°C - +35oC.
#
The surface that your pure water system is to be stood on must be strong enough
to support it (for weight, see "Technical Specifications").
#
A floor drain with waste pipe of DN 50 size (38.5 mm id) is required. Should this not be
available, then a water watcher (article no.: 16.0129) must be installed to protect against
damage from water flooding!
#
An unrestricted gravity flow of concentrate to the floor drain is obligatory.
#
An electrical socket appropriate to the voltage given on the type plate of the system must
be positioned directly alongside the system. The safety fuse must be appropriate for the
power required (see "Technical Specifications").
#
Ensure that there is sufficient working room around the system.
#
An R ¾“R male thread tap water connection which can be shut off must be installed in
the direct vicinity of the system.
#
Easy access to this must be ensured.