LabTower EDI
Ultrapure water system
45
Fault message:
"Measuring cell LF3"
- Cable break at the measur-
ing cell
- System control defect
- EDI conductivity outside of
measuring range
- Replace the measuring cell
- Replace the system control
- see "Lim.val.EDI"
Fault message:
"Measuring cell Temp."
- Cable break at the measur-
ing cell
- System control defect
- Replace the measuring cell
- Replace the system control