background image

16 

 

© Titan Enterprises Ltd 

www.flowmeters.co.uk

  

PulsiteLink_Instruction_Manual_0722 

4.3.3.1.1 

Pin Assignments: 

The options for each PIN are: 

 Not Used  

 Input or 

 Output 

 

Each of the PIN Assignments can be configured as an 
output function or not used 

The operation to modify NPN Output and PNP Output 
is identical within the software, with the exception of the 
type of transistor pulse output. 

 

The options for each PIN, are: 

 Not Used or 

 Output 

NOTE: 

These instructions are the same for NPN and PNP with the exception of the transistor 

wiring and operation. 

For example: For NPN the i

dle state is switch ‘open’ which will be a high V

olt output if there is 

a pull-up in the circuit. PNP would be the opposite, at zero Volt. 

4.3.3.2  NPN Output / PNP Output 

If Output is selected in Pin assignments, the output options are: 

 

Pulse 

 

 

Single or Dual flow alarm 

 

4.3.3.2.1 

Pulse:  

Enter  the  number  of  pulses  per  unit  volume 
required.  This  figure  can  be  adjusted  to  suit  the 
application  and  the  flow  range  required  from  the 
meter.  

The maximum output frequency is 5000 Hz so care 
must be taken to ensure that this pulse rate is not 
exceeded.  

 

NOTE: 

Details of how to connect the outputs can be found in Section 3.1.1 

There are two options for the Pulse Output: 

  

 

 

Asynchronous 

In this mode the meter measures 
according to the analog response time 
and sends pulse (NPN/PNP) output 
after that period at 50:50 ratio. 

 

Synchronous 

In this mode the Pulse output 
(NPN/PNP) is sent directly on receipt of 
the second pulse edge received.  

The pulse output is direct 50:50 ratio or 
can be set to a specific pulse width(ms). 

 

 

Содержание Pulsite Link

Страница 1: ...Titan Enterprises Ltd www flowmeters co uk PulsiteLink_Instruction_Manual_0722...

Страница 2: ...o NPN and PNP The Pulsite Link pulse convertor can also provide o Flow alarm switches via NPN and PNP output terminals Disclaimer This information has been reviewed and believed to be correct at the t...

Страница 3: ...l Failure to follow these guidelines may result in damage to the equipment or personal injury Manufacture Information The Pulsite Link pulse convertor is designed and manufactured by Titan Enterprises...

Страница 4: ...uch as o NPN o PNP o Reed Switch The Pulsite Link pulse convertor will provide linearized outputs of o 4 20mA active o 0 5V o 0 10V The Pulsite Link pulse convertor can also provide o Flow alarm switc...

Страница 5: ...5 Titan Enterprises Ltd www flowmeters co uk PulsiteLink_Instruction_Manual_0722...

Страница 6: ...h max 400Hz VDC Supply to meter Flowmeter 5V Out 0VDC Supply to Flowmeter Flowmeter 0V Pulse from flowmeter Flowmeter Sensor Supply 5Vdc 50mA max room temperature 25mA at high temperature Pull up down...

Страница 7: ...7 Titan Enterprises Ltd www flowmeters co uk PulsiteLink_Instruction_Manual_0722 3 1 1 Wiring Diagrams...

Страница 8: ...8 Titan Enterprises Ltd www flowmeters co uk PulsiteLink_Instruction_Manual_0722...

Страница 9: ...9 Titan Enterprises Ltd www flowmeters co uk PulsiteLink_Instruction_Manual_0722 NOTE Best practice is to place Pull up down Resistor near target device...

Страница 10: ...ng Software screen NOTE When correctly connected the bars will rotate in the bottom left had corner and you should see All Controls enabled On connecting Pulsite Link the Software will normally automa...

Страница 11: ...you wish to connect By opening another copy of the Titan Interface Software you can connect and modify multiple meters from the same computer The panel below shows the main menu function icons of the...

Страница 12: ...shown in the software they can be retrieved from the nonvolatile memory of the flow meter by pressing the looking glass icon 4 1 3 Test Outputs This tab enables the user to manually set outputs on the...

Страница 13: ...eter to present the flow and outputs Flow Details Shows the user real time data from the meter including flow and outputs Multiple Window Display To aid set up whilst operating the meter it is possibl...

Страница 14: ...eter is fundamentally a volumetric device but a density figure can be entered here if one of the mass units is selected Caution must be used however as there is no temperature density correction Cut O...

Страница 15: ...lue will give a recommended lowest flow for the input calibration k factor and the chosen Analog Response Time Setting the cut off value below the recommended level means when the flow is between the...

Страница 16: ...opposite at zero Volt 4 3 3 2 NPN Output PNP Output If Output is selected in Pin assignments the output options are Pulse Single or Dual flow alarm 4 3 3 2 1 Pulse Enter the number of pulses per unit...

Страница 17: ...n can be adjusted with the Invert check box Hysteresis This is used to prevent fluttering of the alarm signal when flow is around the threshold level The value is set around the Threshold value For ex...

Страница 18: ...ggered The remaining settings in the input box are the same as for Single Flow alarms The alarm will automatically activate and reset in proportion to the user set Hysteresis of the Threshold levels T...

Страница 19: ...Powered VDC and ON Unpowered 0VDC Using the PNP OFF Unpowered 0VDC and ON Powered VDC 4 3 3 4 Analog Outputs NOTE FOR FULL ANALOG OUTPUT FUNCTION EXTERNAL POWER OF 15VDC IS REQUIRED Analog Output the...

Страница 20: ...is operating e g relay status and frequency output if these options are selected Parameters Flow This is the instantaneous flow rate and is updated approximately every 100 milliseconds Input The measu...

Страница 21: ...wer 237 Active Output Pulse Frequency Hz Off High Output Alarm State No Alarm VDC On Low Output Alarm State Alarm Triggered 0VDC On Low Output Alarm State Invert Selected No Alarm 0VDC Off High Output...

Страница 22: ...10V 3 22 No Power Analog Voltage output 0V 5V 10V No external power to meter 5 LED Indicators The Pulsite Link has two LED indicators Left Hand Side Right Hand side GREEN Power 6VDC GREEN flashing Pu...

Страница 23: ...Time User Configurable 100ms 1s or 10s Option in software Accuracy 4 20mA output 0 012 linearity 0 15 maximum total output error 16 Bit resolution and monotonicity 0 10 Volt output 0 1 linearity 0 5...

Страница 24: ...he diagrams in the manual No Pulse Output on wire or showing on Titan Interface Software Incorrect settings Check Settings installation No analog output Incorrect Configuration Check that the correct...

Страница 25: ...Connection Return to Titan Enterprises for repair Titan Interface Software is not showing all Tabs User Settings Flow Details Advanced Datalogging Software is placing the tab window outside the monit...

Отзывы: