9
End Node Accelerometer Offset Adjustment:
1. Turn switch unit OFF (LED is not blinking).
2. Turn switch four to the ON position. Remaining switches do not require
adjustment.
3. Turn end node power ON.
4. Within five seconds, position the end node as shown in Figure 8.1
until the LED light flashes.
5. Turn end node power OFF.
6. Turn switch four to the OFF position.
7. While looking at LED light, turn end node power ON. LED light should
flash at least one time.
8. Offset adjustments are complete.
Figure 8.1: StatusCheck End Node
Accelerometer Offset Adjustment
Orientation ———————————
Temperature Probe Description:
The StatusCheck end node has an external spring-loaded temperature
probe that measures the surface temperature of the machinery to which it
is attached. The RTD element, positioned at the probe tip, contacts the
machine for quick temperature measurement.
The temperature probe is available in three different versions:
S = Standard (magnetic flush mount or with use of tack weld plate)
L = Long Probe (for use with threaded nipples or for internal
temperature measurements)
CXX = Custom (C followed by exact length from uncompressed tip to
bottom of flange)
Note:
The custom length option gives customers the ability to extend the
range of the probe to contact internal interfaces of a monitored machine.
This feature can be used to obtain bearing cup O.D. temperatures as well
as other important internal temperature values. Temperature probe length
is specified when placing a StatusCheck end node order.
Содержание StatusCheck
Страница 1: ...StatusCheck Operators Manual StatusCheck 2 4GHz ...
Страница 23: ......