7 Alarms and messages
50
IFU, MIRUS Controller, en-GB, N.00
7.1.5
Technical alarms
Alarm
(in alphabetical
order)
Priority
Device
action
Set condition
Reset condition by
user
Device
inclined.
Low
Alarm and
stop VA
delivery.
Tilt angle > 25° for
t < 30 sec.
Reposition device on
level surface.
High
Alarm and
stop VA
delivery.
Tilt angle > 25° for
t ≥ 30 sec.
Dosage
failed!
Start VA?
High
Alarm and
stop VA
delivery.
Calculation problem
was detected.
Restart VA delivery by
pressing Play and
confirm or turn off
device.
Failure! Turn
off.
High
Alarm and
stop VA
delivery.
A technical failure was
detected.
Turn off device.
Fill port flap
still open.
Low
Alarm
Fill port flap open
>5 min.
Close fill port flap.
High
Alarm
Fill port flap open
>7 min.
Heating
system failed.
High
Alarm
No activity of the
heating system could
be detected.
Turn off device.
Interface
disconnected.
High
Alarm and
stop VA
delivery.
Interface plug
disconnected while in
operation.
Reconnect interface
plug.
Mains supply
lost.
Low
Alarm
AC lost.
Connect mains supply.
Occlusion of
Interface
High
Alarm and
stop VA
delivery.
A gas sampling or
measurement line is
clocked.
Check interface for
kinking.
Check fresh gas flow of
AWS to be min 1.5 *
MV.
Overtem-
perature
Low
Alarm
Inner Controller
temperature > 55°C
Check fan input and
output.
Pause still
active.
Low
Alarm and
pause VA
delivery.
Delivery paused for >
2 min.
Neutralise interception
by pressing Pause.
Please
replace
Reflector
promptly!
Low
Alarm
No valid values from
Reflector.
Replace reflector.
(System test should be
repeated as soon as
possible.)
Reflector
expired.
Low
Alarm
Interface not released
for >160 hours.
Prepare to replace or
replace reflector.
Reflector
soon expires.
Low
Alarm
MIRUS Interface not
released for 144
–
160 hours.
Prepare to replace or
replace reflector.
Reservoir
needs refill
Low
Alarm
Internal reservoir fill
level < 60 mL
Refill reservoir.
Reservoir
empty
High
Alarm
Internal reservoir fill
level < 45 mL.
Refill reservoir.