6 Maintenance and Cleaning
IFU, MIRUS Controller, en-GB, N.00
43
6
Maintenance and Cleaning
6.1
Maintenance
6.1.1
General Information
Maintenance, safety checks and maintenance measures must be performed only by the
manufacturer or by qualified personnel authorized by the manufacturer.
6.1.2
Schedule
Affected part
Period
Performer
MIRUS Controller
Every 12 months Safety Check
Visual Inspection
Electrical Safety Test
Checking gas monitor
Checking pump
Safety Function test
Manufacturer or qualified
personnel specially
authorized by the
manufacturer
UPS Battery pack
When not using the device, leave
device connected to mains supply or
connect to mains supply at least once
a month for at least 24h.
Replace every 2 years.
Manufacturer or qualified
personnel specially
authorized by the
manufacturer
Fan filter
Replace every 12 months.
Manufacturer or qualified
personnel specially
authorized by the
manufacturer
VA reservoir
Empty every 4 weeks.
Do not re-use VA.
User / Operator
Software
Manufacturer gives information as
soon as software update is available.
Manufacturer or qualified
personnel specially
authorized by the
manufacturer