1 Introduction
6
IFU, MIRUS Controller, en-GB, N.00
1
Introduction
1.1
Intended use
The MIRUS Controller is part of the MIRUS System. It is intended to be used for the application
of volatile anaesthetic agents to humans with a tidal volume of
≥ 200 mL.
1.2
Intended purpose
The MIRUS Controller is designed to work only in combination with the MIRUS Reflector. For
hygiene safety reasons connection of the MIRUS System to a patient is only possible via the
MIRUS Filter.
Note:
For more information about the MIRUS Reflector and the MIRUS Filter please refer to
the manufacturer’s IFU.
1.3
C
ontraindication
s
Do not use the MIRUS System with
patients requiring less than 200 mL tidal volume.
patients that have a contraindication for the application of volatile anaesthetic
agents.
pure spontaneously breathing patients without a ventilator securing apnoea
back up ventilation mode.
jet or high frequency ventilation.
leaking patient airway systems, such as face mask or helmet systems.
additional filters (HME/F) or tubes that significantly increase dead space.
active humidification.
low flow anaesthesia.
1.4
MIRUS System
The MIRUS System consists of three devices:
MIRUS Controller (MC)
The electrical device that contains the anaesthetic agent and controls the agent
delivery.
MIRUS Reflector (MR)
The multi patient device connected to the MIRUS Controller, inserted into the
patient’s breathing system.
MIRUS Filter (MF)
The single patient device connected to the MIRUS Reflector, protecting the patient
and breathing system.