2 Safety
IFU, MIRUS Controller, en-GB, N.00
17
2.5.5
Electromagnetic compatibility
CAUTION
Do not use the MIRUS System in the presence of an MRI System.
WARNING
The use of other electrical equipment, e.g. the line supply cable, may
cause a higher EM-emission or may weaken the immunity of the
device. This may cause a risk to the patient.
The use of other electrical equipment on or near this system may cause
interference. Verify normal operation of equipment in your
configuration before connecting a patient.
2.6
Residual risk
– Fail-safe mode
During its operation, the MIRUS Controller is monitored by a separate safety system. In case
of a failure, the safety system shifts the device to a fail-safe mode, which is safe for the patient.
During the fail-safe mode there is no delivery of anaesthetic to the patient.
Safe mode means:
screen is blank
no volatile anaesthetic agent is
delivered
alarm lights are flashing in red
confirm button is flashing in red
audible alarm sound
To silence the audible alarm, press the Confirm button for at least 4 seconds.
MIRUS
Controller now turns off flashing red LEDs and stops alarm sound. To turn off the fail-safe
mode disconnect mains supply and press the Confirm button again for at least 4 seconds. The
red LEDs now turn off.
To restart MIRUS Controller after a fail-safe mode
wait for 5 seconds,
reconnect AC supply,
re-start by pressing the Confirm button (usual start).