6 Maintenance and Cleaning
46
IFU, MIRUS Controller, en-GB, N.00
WARNING
Ensure that the housing and the touch screen have dried completely
before reconnecting the MIRUS Controller to an electrical supply.
MIRUS Controller Fill port
Always keep the fill port flap closed to avoid the entry of dust or other substances. In case
the fill port needs to be cleaned use a clean and lint free cloth. Avoid using cotton swabs as
they too will leave small particles and lint in the fill port.
MIRUS Reflector
To clean the external surfaces of the MIRUS Reflector follow the instructions for use supplied
with the product.
MIRUS Filter
The MIRUS Filter is a single use device and should not be soaked, rinsed, washed, sterilised
or treated with liquid disinfectants.
6.3
Draining the reservoir
To drain the reservoir, a draining kit from the manufacturer TIM is needed (refer to chapter
9
Parts list
for ordering information).
Perform the draining according to the manufacturer’s
guide that comes with the draining kit.
6.4
Shipping MIRUS Controller
If the MIRUS Controller is to be sent, it must be cleaned and disinfected (refer to chapter
6.2 Cleaning
), as well as completely emptied (refer to chapter
6.3 Draining the reservoir
).
Only use the original packing material to pack the MIRUS Controller:
Safety bag
Transport jacket (top and bottom frame)
MIRUS Controller box white
MIRUS Controller shipping box, 2-piece
If the original packing material is not available, it can be ordered via:
Perform the packing as describ
ed in the manufacturer’s guide.
Содержание MIRUS Controller