background image

20

2.3 Product Description

The VSR vacuum transducer is measuring total gas pressure in the range of
5.0x10

-5

- 2000 mbar. The transducer can be connected to Thyracont display and

control units or to customer related power supply and evaluation units in complian-
ce with pin assignment.
The analog output signal 1.2 V - 8.8 V has a logarithmic dependence on pressure
over the whole range.
The VSR is equipped with a metal-sealed combination sensor type Piezo / Pira-
ni and temperature compensated. It can be mounted to suitable flange connectors.

Proper Use

The VSR serves exclusively to provide absolute pressure measurements in ga-
seous media. It may only be connected to components specifically provided for
such purpose.
The device is classified in electromagnetic interference class A and therefore can
cause radio interference in living quarters.

Improper Use

The use for purposes not covered above is regarded as improper, in particular:

connection to components which are not specified for in their operating in-
structions

connection to components containing touchable, voltage carrying parts.

No liability or warranty will be accepted for claims arising from improper use.

The user bears the responsibility with respect to the used process media.

The device is not designed for use in corrosive gas atmosphere.
Aggressive media such as halogenides, carbon or oxygen plasma
can reduce sensor life-time!
Dust, oil or condensing vapours will affect sensor performance and
may cause malfunction!

VSR-0-10 V-210805

Содержание VSR53MV

Страница 1: ...V a c u u m I n s t r u m e n t s VSR53MV VSR54MV Vakuum Transmitter Vacuum Transducer Betriebsanleitung Operating Instructions...

Страница 2: ......

Страница 3: ...ktrischer Anschluss 8 3 3 1 Anschluss an Thyracont Anzeigeger te 8 3 3 2 Kundeneigene Spannungsversorgung 9 4 Betrieb 10 4 1 Allgemeines 10 4 2 Nachjustieren 11 5 Wartung und Service 12 6 Technische D...

Страница 4: ...d Schutzma nahmen Ber cksichtigen Sie m gliche Reaktionen zwischen Werkstoffen und Prozessmedien z B infolge der Eigenerw rmung des Produkts Ger t nicht eigenm chtig umbauen oder ver ndern Informieren...

Страница 5: ...R Staubschutzkappe Betriebsanleitung Lieferbares Zubeh r Zentrierdichtring DN16KF mit Metall Drahtfilter ZZDF016 Zentrierdichtring DN16KF mit Metall Schutzblende ZZCH016 Stecker und Messkabel Messkabe...

Страница 6: ...Komponenten angeschlossen werden Der Messumformer ist ein Ger t der St rklasse A und kann daher im Wohnbereich zu Funkst rungen f hren Nicht bestimmungsgem e Verwendung Als nicht bestimmungsgem gilt...

Страница 7: ...recht ber KF Kleinflansch CF Flansch herstellen F r Kleinflansch Verbindung Metall Spannelemente verwenden die sich nur mit einem Werkzeug ffnen und schlie en lassen Dichtringe mit Zentrierring verwen...

Страница 8: ...hr standhalten Dies kann zu Gesundheitssch den durch ausstr mende Prozessmedien f hren 3 3 Elektrischer Anschluss 3 3 1 Anschluss an Thyracont Anzeigeger te Wird der Transmitter an einem Thyracont Anz...

Страница 9: ...gem untenstehender Pinbelegung herzustellen Stecker Hirschmann 6polig m nnlich 3 1 2 4 5 6 Pin1 Identifikation 12 k Pin2 Signal Output 0 10 VDC Pin3 AGND Pin4 Voltage Supply 12 30 VDC Pin5 Supply GND...

Страница 10: ...Absolutdruck Ausgangssignal Das Mess Signal 1 2 V 8 8 V des VSR ist ber den gesamten Messbereich von 5 0x10 5 2000 mbar logarithmisch vom Druck abh ngig Die Umrechnung erfolgt gem folgendem Zusammenha...

Страница 11: ...Druck kleiner 5 0x10 6 mbar sein In allen anderen F llen wird der Istwert eines Referenz Druckmessger ts ben tigt Zum Justieren Gummistopfen ber den Tastern entfernen 1 und mit einem d n nen Schraube...

Страница 12: ...n chste Seite zur Reparatur an uns Das Ger t ist nicht zur kundenseitigen Reparatur vorgesehen Fehlfunktionen des Ger tes die auf Verschmutzung oder Ver schlei zur ckzuf hren sind fallen nicht unter...

Страница 13: ...13 VSR 0 10 V 210805...

Страница 14: ...ar 2 v Messwert Materialien mit Vakuumkontakt Edelstahl 1 4307 Wolfram Nickel Glas SiO2 Zinn Silber Lot Polyimid Epoxid Reaktionszeit 18 ms Betriebstemperatur 5 60 C Lagertemperatur 40 65 C Ausheiztem...

Страница 15: ...15 Schutzart IP 40 Gewicht ca 100 g VSR53 VSR 0 10 V 210805...

Страница 16: ...16 7 Konformit tserkl rung VSR 0 10 V 210805...

Страница 17: ...on 22 3 3 1 Operation with Thyracont Display Unit 22 3 3 2 Operation with other Supply and Evaluation Units 23 4 Operation 24 4 1 General 24 4 2 Readjustment 25 5 Maintenance and Service 26 6 Technica...

Страница 18: ...o cess media used Consider possible reactions between materials and process media e g due to the heat generated by the product Do not carry out any unauthorized conversions or modifications on the uni...

Страница 19: ...y consignment are Transducer VSR Protective flange cover Operating instructions Available Accessories Centering ring DN16KF with metal wire filter ZZDF016 Centering ring DN16KF with metal baffle ZZCH0...

Страница 20: ...e connected to components specifically provided for such purpose The device is classified in electromagnetic interference class A and therefore can cause radio interference in living quarters Improper...

Страница 21: ...se clamps that can be opened and closed with appropriate tools only use sealing rings with a centering ring Make sure that the sensor flange is connected to ground e g by having electrical contact to...

Страница 22: ...Unit For operation of the transducer with a Thyracont display and control unit a suitable measurement cable must be used see accessories Do not connect or disconnect the transducer when the cable is o...

Страница 23: ...s considering EMI demands and according to the pin description shown below Socket Hirschmann 6pole male 3 1 2 4 5 6 Pin1 Identification 12 k Pin2 Signal Output 0 10 VDC Pin3 AGND Pin4 Voltage Supply 1...

Страница 24: ...arithmic dependence on pressure over the whole measurement range 5 0x10 5 2000 mbar Conversion of voltage signal and pressure is done according to the following formula Vout V log p mbar 5 5 p mbar 10...

Страница 25: ...be less than 5 0x10 6 mbar In all other cases a reference pressure reading of a standard gauge will be required For adjustment first remove the rubber caps from both pushbuttons 1 Then by means of a s...

Страница 26: ...ration of contamination see next page The unit is not prepared for customer repair Malfunction of the unit which is caused by contamination or wear and tear is not covered by warranty Error Messages a...

Страница 27: ...27 VSR 0 10 V 210805...

Страница 28: ...mbar 2 f reading Materials in contact with vacuum stainl steel 1 4307 tungsten nickel glass SiO2 tin silver solder polyimide epoxy Reaction time 18 ms Operating temperature 5 60 C Storage temperature...

Страница 29: ...29 Protection class IP 40 Weight approx 100 g VSR53 VSR 0 10 V 210805...

Страница 30: ...30 7 Declaration of Conformity VSR 0 10 V 210805...

Страница 31: ...VSR 0 10 V 210805...

Страница 32: ...VSR 0 10 V 210805...