23
3.3.2 Operation with other Supply and Evaluation Units
The transducer can be operated with other customer related display units or volta-
ge supplies.
The electrical connection is to be made by means of suitable cables considering
EMI demands and according to the pin description shown below:
Socket
Hirschmann, 6pole, male
3
1
2
4
5
6
Pin1:
Identification: 12 k
Ω
Pin2:
Signal Output 0-10 VDC
Pin3:
AGND
Pin4:
Voltage Supply 12...30 VDC
Pin5:
Supply GND
Pin6:
Shield
We recommend to have Shield (pin 6) and supply common (pin 5)
grounded in the supply unit.
Incorrect connection or inadmissible supply voltage can damage
the transducer!