30
Overpressure in the vacuum system > 1 bar
Accidental or unintended opening of clamp elements under stress
can lead to injuries due to parts flying around! Unsecured hose
connections can release, process media thus can leak and possibly
damage your health!
Overpressure in the vacuum system 1.5 to 4 bar
KF flange connections with elastomer sealings cannot withstand
such pressures. Process media thus can leak and possibly damage
your health!
3.3 Electrical Connection
3.3.1 Voltage Supply / RS485 Interface
The electrical connection is to be made by means of suitable cables considering
EMI demands and according to the pin description shown below:
Socket
M12, A-coded, 5-pole, female
Pin1:
RS485 +
Pin2:
Voltage Supply 24 VDC
Pin3:
Supply GND
Pin4:
RS485 -
Pin5:
n.c.
We recommend to have shield and supply common (pin 3) groun-
ded in the supply unit.
Incorrect connection or inadmissible supply voltage can damage
the transducer!
VSH-EtherCAT-200116