7-
53
SAVE
SAVE
=
password
saves calibration coefficients, printer configuration, and the
pressure to non-volatile memory, overwriting any existing
values. The temperature calibration coefficients only affect the
active probe (see
PROB
e command). The
password
, 5A1MP,
is required in order to prevent accidental alteration of non-
volatile memory.
________________________________________________________________________
IBM BASICA Example (assuming COM1 initialized to #1 and 5A-1MP
ECHO
is OFF):
.
.
.
100
PRINT #1, "PROB=1"
'Set Probe 1
110
PRINT #1, "DZERO?"
'Get dry bulb zero
120
INPUT #1, A
'Coefficient
130
PRINT #1,"DZER=";A + 0.25
'Offset dry bulb by .25•C
140
PRINT #1, "SAVE-5A1MP:
'
SAVE
to non-volatile memory
.
.
.
________________________________________________________________________
TERMINAL Example (assuming
ECHO
is ON):
PRES = 24.50
Set to 24.50 "Hg
SAVE = 5A1MP
Save to non-volatile memory
________________________________________________________________________
IEEE 488 Example (HP Rocky Mountain BASIC syntax):
.
.
.
100
OUTPUT 705; "DOZE=401.640;
DOSP=.33323116"
!Change Ohms coefficients.
110
OUTPUT 705; "DOLI=.000082441619;
SAVE=5A1MP:
!Save them.
______________________________________________________________________________________
RELATED COMMANDS:
REST
ore
PROB
e
Содержание 5A-1MP
Страница 3: ...5A 1MP Laboratory Reference Psychrometer...
Страница 45: ...7 18 110 ENTER 705 X 120 DISP X Display it RELATED COMMANDS DSPAn DZERo PROBE RESTore SAVE...
Страница 67: ...7 40 ________________________________________________________________________ RELATED COMMANDS AUTO PRINt...
Страница 71: ...7 44 ________________________________________________________________________ RELATED COMMANDS RESTore SAVE...
Страница 83: ...7 56 ________________________________________________________________________ RELATED COMMANDS DATE...