4 - 1
Section 4
MAINTENANCE
4.1 PROBE WICK CHANGEOUT INSTRUCTIONS
The CK-1 Wick Changeout Kit is provided to easily change wicks in the sensing probe.
Wick changes should be performed after approximately 40 hours of operation, or more
often, particularly if the instrument is in a dirty environment. A replacement wick supply
should be kept on hand and can be ordered from the factory in advance of the need.
Consult the factory for current price information in the event a replacement supply is
needed.
To change the wick, perform the following:
1) Place probe upside down upon two blocks of wood or other material, one on each
side of the probe handle.
2) Remove reservoir plug, cover screw, and rubber feet.
3) Gently pry reservoir cover with plastic tool and remove.
4) Carefully remove the four housing retainer screws located in the corners of the probe;
they join the two halves of the probe.
5) Gently remove bottom half of probe housing, feeding the wick through the vertical
teflon wick tube. Keep the large horizontal teflon air sleeve in place during removal.
Note
- Blower motor is also loose once housing is removed.
6) Lift teflon air sleeve up and off the sensor mounts taking care to feed the wick
through the air sleeve avoiding any lateral movement and undue strain upon the sensor.
7) The sensor and wick are now in the open where the wick may be removed. Using the
dropper provided, apply several drops of ethyl alcohol to the portion of the wick
covering the sensor. Allow to soak for approximately one minute, replenishing the
alcohol as evaporation occurs.
8) Using one of the picks provided, gently probe the wick covering the sensor to loosen.
9) Using the curved tweezers provided, very gently push the wick up on itself or up on
the sensor, assisting with the other hand from the reservoir side. (The wick construction
is very similar to a "Chinese finger puzzle" in that pushing tends to release it where
pulling only tightens the wick and compounds the problem of removal.) The wick
should expand. When this occurs, use the tweezers to slide the wick from the sensor.
Содержание 5A-1MP
Страница 3: ...5A 1MP Laboratory Reference Psychrometer...
Страница 45: ...7 18 110 ENTER 705 X 120 DISP X Display it RELATED COMMANDS DSPAn DZERo PROBE RESTore SAVE...
Страница 67: ...7 40 ________________________________________________________________________ RELATED COMMANDS AUTO PRINt...
Страница 71: ...7 44 ________________________________________________________________________ RELATED COMMANDS RESTore SAVE...
Страница 83: ...7 56 ________________________________________________________________________ RELATED COMMANDS DATE...